User's Manual

Table Of Contents
Page 34 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL 7ML19985KB01
mmmmm
Quick Start: local
Operation
Low Calibration Point
High Calibration Point
Wizard Complete
To transfer Quick Start values to the device and return to Program menu, press
DOWN arrow (Finish). Then press LEFT arrow to return to Measurement mode.
Operation Description
LEVEL (1) *
Distance from Low Calibration Point to material surface
SPACE (2)
Distance from High Calibration Point to material surface
DISTANCE(3)
Distance from Sensor Reference Point to material surface
Options
BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wizard menu when
Quick Start is successfully completed.)
High
Calibration
Point
Low
Calibration
Point
level
space
distance
sensor reference point
(flange face: see
Dimensions
on page
11)
Distance from Sensor Reference Point to
Low Calibration Point: usually process
empty level.
Values
Range: 0.0000 to 100.000 m
Distance from Sensor Reference Point to
High Calibration Point: usually process full
level.
Values
Range: 0.0000 to 100.000 m