B: Troubleshooting
Display shows
S: 0 LOE
location or aiming:
poor installation
Auto False Echo Suppres-
sion may be incorrectly
applied
ensure antenna has clear view of mate-
rial to be measured
review “Shaper Mode (2.8.4.)” on
page 126
ensure Auto False Echo Suppression
Range is set correctly
Display shows
S: 0 LOE
antenna malfunction:
temperature too high
•physical damage
Check for physical damage and replace
sensor if required.
Reading does not
change, but the level
does
SITRANS LR560 processing
wrong echo, for example, ves-
sel wall, or structural member
re-locate SITRANS LR560
check nozzle for internal burrs or welds
•use Shaper Mode (2.8.4.) if necessary:
see “Shaper Mode (2.8.4.)” on page 126
Measurement is consis-
tently off by a constant
amount
•setting for Low Calibration
Pt. (2.3.1.) not correct
•setting for Sensor Offset
(2.3.3.) not correct
check distance from sensor reference
point to Low Calibration Pt. (2.3.1.)
•check Sensor Offset (2.3.3.)
Screen blank
power error
check device label rating against voltage
supply
check power wiring or source
too much load resistance
remove something from the loop, or
increase supply voltage
Reading erratic
echo confidence weak
•refer to Confidence (2.7.6.1.)
•use Shaper Mode (2.8.4.) and Shaper
Mode (2.8.4.)
material filling re-locate SITRANS LR560
Reading response slow
Fill Rate/Min (2.4.2.) setting
incorrect
increase measurement response if possi-
ble
Reads correctly but occa-
sionally reads high when
vessel is not full
detecting close range
echo
build up near top of
vessel or nozzle
nozzle problem
Use the air purge feature to clean the
antenna
•use Shaper Mode (2.8.4.) and Shaper
Mode (2.8.4.)
Level reading lower than
actual material level
material is within Near
Range zone
multiple echoes
processed
decrease Near Range (2.7.1.) (minimum
value depends on antenna type)
raise SITRANS LR560
ensure Algorithm (2.7.3.1.) is set to F (First
echo)
vessel near empty and
low dK material
•set CLEF Range (2.7.3.4.) to 0.5 m
Operation Troubleshooting (Continued)
Symptom Cause Action