User's Manual

Table Of Contents
Page 118 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL 7ML19985KB01
mmmmm
B: Troubleshooting
S: 29
* EEPROM damaged.
Repair required: contact your local Sie-
mens representative.
S: 30 EEPROM corrupt. Reset power. If fault persists, contact
your local Siemens representative.
S: 31
*Flash error.
Repair required: contact your local Sie-
mens representative.
S: 33
*
Factory calibration for the internal temper-
ature sensor has been lost.
Repair required: contact your local Sie-
mens representative.
S: 34
*
Factory calibration for the device has been
lost.
Repair required: contact your local Sie-
mens representative.
S:39
* Temperature sensor failure.
Repair required: contact your local
Siemens representative.
S:40
Temperature too high.
Relocate device and/or lower process
temperature enough to cool device. Inspect
for heat-related damage and contact your
local Siemens representative if repair is
required.
S:41
Temperature too low.
Relocate device and/or raise process
temperature enough to warm device.
Inspect for temperature-related damage
and contact your local Siemens
representative if repair is required.
S: 48
*
User configuration is invalid. One or more
of parameters: Low Calibration Point, High
Calibration Point, Volume breakpoints, and/
or Auto False-Echo Suppression, are set to
invalid values.
Reconfigure the unit.
Ensure the difference between High Cali-
bration Point and Low Calibration Point is
not less than or equal to zero; do a Mas-
ter Reset.
General Fault Codes (Continued)
Code /
Icon
Meaning Corrective Action