User's Manual
Table Of Contents
- Safety Notes
- FCC Conformity
- CE Electromagnetic Compatibility (EMC) Conformity
- Industry Canada
- The Manual
- Technical Support
- SITRANS LR560 Overview
- Specifications
- Installation
- Wiring
- Local operation
- Operating via SIMATIC PDM
- Operating via FDT (Field Device Tool)
- Operating via AMS Device Manager
- AMS Menu Structure
- Parameter Reference
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Appendix F: HART Communications
- Appendix G: Firmware Revision History
- Glossary
- Index
- LCD menu structure
Page 116 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL 7ML19985KB01
mmmmm
B: Troubleshooting
General Fault Codes
• Data exchanged
• No data exchange
•Write access enabled
•Write access disabled
Notes:
• If more than one fault is present, the device status indicator and text for each fault
alternate at 2 second intervals.
• Some faults cause the device to go to Fail-safe mode (Fault 52). These are indicated with
an asterisk (*).
General Fault Codes
Code /
Icon
Meaning Corrective Action
S: 0
*
The device was unable to get a measure-
ment within the Fail-safe LOE Timer period.
Possible causes: faulty installation,
antenna material buildup, foaming/other
adverse process conditions, invalid calibra-
tion range.
• Ensure installation details are correct.
• Ensure no antenna material buildup.
Clean if necessary.
• Adjust process conditions to minimize
adverse conditions.
• Correct range calibration.
• If fault persists, contact your local Sie-
mens representative.
S: 2
*
Unable to collect profile because of a
power condition that is outside the operat-
ing range of the device.
Repair required: contact your local Sie-
mens representative.
S: 3
Device is nearing its lifetime limit according
to the value set in Maintenance Required
Limit.
Replacement is recommended.
Icon
Priority
Level
Meaning (Continued)