7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 115
mmmmm
B: Troubleshooting
Device Status Icons
Icon
Priority
Level
Meaning
1
• Maintenance alarm
• Measurement values are not valid
2
• Maintenance warning: maintenance demanded immediately
• Measured signal still valid
3
• Maintenance required
• Measured signal still valid
1
• Process value has reached an alarm limit
2
• Process value has reached a warning limit
3
• Process value has reached a tolerance limit
1
•Configuration error
• Device will not work because one or more parameters/components is
incorrectly configured
2
•Configuration warning
• Device can work but one or more parameters/components is incorrectly
configured
3
• Configuration changed
• Device parameterization not consistent with parameterization in project.
Look for info text.
1
• Manual operation (local override)
• Communication is good; device is in manual mode.
2
• Simulation or substitute value
• Communication is good; device is in simulation mode or works with substi-
tute values.
3
• Out of operation
• Communication is good; device is out of action.