Parameters
4.8.7. Maintenance Status
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Calibration Schedule Status window.
4.8.8. Acknowledge Status
Indicates which level of maintenance reminder has been acknowledged.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Calibration Schedule Status window.
4.8.9. Acknowledge
Acknowledges the current maintenance reminder.
To acknowledge a reminder via SIMATIC PDM:
a) Open the menu View – Device Status and click on the Maintenance tab.
b) In the Calibration Schedule Status section click on Acknowledge
Warnings.
To acknowledge a reminder via the the local control buttons or handheld
programmer:
a) Press RIGHT arrow twice to open parameter view and activate Edit Mode.
b) Press RIGHT arrow to acknowledge the reminder.
5. Communication
5.1. Device Address
Sets the device address or poll ID on a HART network. Any address other than 0 will
cause the output current to be a fixed value, and the current will not indicate the
reading. Range should be limited to 0 to 15 for a HART 5 master.
To reset Device Address via SIMATIC PDM:
Open the project in Process Device Network View then right-click on the
device.
•Go to Object Properties > Connection to access the field Short Address.
Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.
Values
Range:
0 to 63
Default: 0
Note: SIMATIC PDM only allows addresses 0-15.