Page 108 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL 7ML19985KB01
mmmmm
Parameters
4.8.3. Reminder 1 (Required)
If Time until Next Calibration (4.8.2.) is equal to or less than this value, the device
generates a Maintenance Required reminder.
a) Modify values as required.
b) Set Reminder Activation (4.8.5.) to the desired option.
4.8.4. Reminder 2 (Demanded)
If Time until Next Calibration (4.8.2.) is equal to or less than this value, the device
generates a Maintenance Demanded reminder.
a) Modify values as required.
b) Set Reminder Activation (4.8.5.) to the desired option.
4.8.5. Reminder Activation
Allows you to enable a maintenance reminder.
a) First set the values in Reminder 1 (Required) (4.8.3.)/Reminder 2
(Demanded) (4.8.4.).
b) Select the desired Reminder Activation option.
4.8.6. Calibration Interval
User-configurable recommended time between product calibrations.
Values
Range: 0 to 20 years
Default: 0.164 years
Values
Range: 0 to 20 years
Default: 0.019 years
Note: To modify this parameter via SIMATIC PDM it must be accessed via the
pull-down menu Device – Maintenance.
Values
*Timer OFF
ON
NO LIMITS
ON REMinder 1 (Maintenance Required) checked
ON
REMinders 1 and 2 checked
ON
REMinder 2 (Maintenance Demanded) checked
Note: The device always operates in years. Changing the units affects only
the parameter view of the Calibration Interval parameters in SIMATIC PDM.
Values
Units
a)
: hours, days, years
a)
Units are selectable only via SIMATIC PDM.
Range: 0 to 20 years
Default: 1.0 year