7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 99
mmmmm
Parameters
3.6.3. Reminder 1 (Required)
If Remaining Lifetime (3.6.2.) is equal to or less than this value, the device
generates a Maintenance Required reminder.
a) Modify values as required.
b) Set Reminder Activation (3.6.5.) to the desired option.
3.6.4. Reminder 2 (Demanded)
If Remaining Lifetime (3.6.2.) is equal to or less than this value, the device
generates a Maintenance Demanded reminder.
a) Modify values as required.
b) Set Reminder Activation (3.6.5.) to the desired option.
3.6.5. Reminder Activation
Allows you to enable a maintenance reminder.
3.6.6. Lifetime Expected
Allows you to override the factory default.
Values
Range: 0 to 20 years
Default: 0.164 years
Values
Range: 0 to 20 years
Default: 0.019 years
Note: To modify this parameter via SIMATIC PDM it must be accessed via the
pull-down menu Device – Maintenance.
Options
Reminder 1 (Maintenance Required)
Reminder 2 (Maintenance Demanded)
Reminders 1 and 2
*OFF
Note: The device always operates in years. Changing the units affects only
the parameter view of the Remaining Device Lifetime parameters in
SIMATIC PDM.
Values
Units
a)
: hours, days, years
a)
Units are selectable only via SIMATIC PDM.
Range: 0 to 20 years
Default: 10.00 years