7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 95
mmmmm
Parameters
To use TVT shaper via LDI (local display interface):
a) Go to Shaper Mode (2.8.4.) and select On.
b) Go to Breakpoint 1-9 (2.9.1.).
c) Open Shaper 1 and enter the TVT Offset value (between –50 and 50).
d) Go to the next Shaper point and repeat steps c) and d) till all desired
breakpoint values have been entered.
2.9.1. Breakpoint 1-9
2.9.2. Breakpoint 10-18
2.9.3. Breakpoint 19-27
2.9.4. Breakpoint 28-36
2.9.5. Breakpoint 37-45
2.9.6. Breakpoint 46-54
2.9.7. Breakpoint 55-63
2.9.8. Breakpoint 64-72
2.9.9. Breakpoint 73-81
2.9.10. Breakpoint 82-90
2.9.11. Breakpoint 91-99
2.9.12. Breakpoint 100-108
2.9.13. Breakpoint 109-117
2.9.14. Breakpoint 118-120
2.10. Measured Values
Read only. Allows you to view measured values for diagnostic purposes.
To access measured values via SIMATIC PDM:
Open the menu View – Process Variables.
2.10.1. Main Output
The value for Level/Volume, always displayed here in percent.
2.10.2. Ouput No Linearization
The value for Level, always displayed here in percent.
2.10.3. Output No Offsets
The value for distance, displayed in sensor units.