2.5.6. TVT shaper
A breakpoint on the TVT curve, normalized to 0.
2.5.6.1. Shaper A (1-9)
2.5.6.2. Shaper B (10-18)
2.5.6.3. Shaper C (19-27)
2.5.6.4. Shaper D (28-36)
2.5.6.5. Shaper E (37 - 40)
2.5.7. Rate
2.5.7.1. Fill Rate
Defines the maximum rate at which the reported sensor value is allowed
to increase.Allows you to further adjust the LR 460 response to increases
in the actual material level (or an advance to a higher Failsafe Material
Level, P071). Fill Rate is automatically updated whenever Response Rate
is altered.
Enter a value slightly greater than the maximum vessel-filling rate, in
Sensor Units per minute.
Values
Range: 0.0000 to 99999 m / min.
Factory setting: 0.1
Altered by Response Rate
Related Unit (Level)
High Level Point
Handheld programmer Value Meters/Minute
10.1