User's Manual
Table Of Contents
7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 72
mmmmm
Parameters (Profibus PA)
4.5.4.2: Flange type Type of flange according to
customer
specifica-
tions
DIN
ANSI
JIS
Special design
4.5.4.3: Pressure
range
Pressure range of the
process connection
according to
customer
specifica-
tions
non-editable
4.5.4.4: Antenna type according to
customer
specifica-
tions
Horn type 2"
Horn type 3"
Horn type 4"
4.5.4.5: Antenna
extension
For qualified personnel only
4.5.4.6: Flange
material
according to
customer
specifica-
tions
304
316/316L
4.5.4.7: Seal material Sealing material according to
customer
specifica-
tions
PTFE
FKM
4.5.5: Tag Device identification according to
customer
specifica-
tions
up to any eight
characters
4.5.6: Descriptor Measuring point descrip-
tion
according to
customer
specifica-
tions
up to any 16 char-
acters
4.5.7: Message Measuring point mes-
sage, e.g. the date of last
check or clean
according to
customer
specifica-
tions
up to any 32 char-
acters
4.5.8: Manufacturer
identification
4.5.8.1: Serial no. Factory serial number unique num-
ber
non-editable
Parameters (PROFIBUS PA)
Note: For parameter descriptions, please refer to the associated HART parameters on
pages 30 to 58.
Parameter, menu
identification
Description
Factory
Setting
Setting
Possibilities