Installation
Electrical Connection
AC version:
The equipment must be protected by a 15A fuse or circuit breaker in the building
installation.
A circuit breaker or switch in the building installation, marked as the disconnect
switch, shall be in close proximity to the equipment and within easy reach of the
operator.
All field wiring must have insulation suitable for at least 250 V.
DC version:
DC input terminals shall be supplied from an SELV source in accordance with IEC
1010-1 Annex H.
Notes (AC and DC versions):
4-20 mA, PROFIBUS PA, DC input circuits, 14 - 20 AWG, shielded copper
wire
AC input circuit, min 14 AWG copper wire
Recommended torque on terminal clamping screws, 0.5 - 0.6 Nm.
1264
5
378
L
1
L
2
N
mA
Rated temperature of
connection cables must
exceed m aximum ambient
temperature by at least 15 K