Specifications
Specifications
SITRANS LR 400
Power
• Power Supply 120 to 230 V AC, ±15%, 50/60 Hz, 6W (12VA) or
24 V DC, +25/-20%, 6W
Power failure: bridge of at least 1 mains period (> 20 ms)
Fuse
• Fuse (both AC and DC versions)
SI1 Fast acting ceramic, 4 x 20 mm, 1A, 250 V AC
SI2 Slow-Blow, 4 x 20 mm, 0.63 A, 250 V AC
Performance
Measured value error (under reference conditions)
• Measuring error ≤ ± 5 mm at 1 to 10 m distance
≤
± 15 mm at 10 to 45 m distance
• Dead zone 0 to 350 mm from bottom edge of flange
• Additional contribution of ≤ 0.1% of the measured value
analog output ≤ 0.05 % of full scale
• Long-term stability ≤
± 1 mm/year
• Repetitive accuracy ≤
± 1 mm at 0 to 45 m, damping ≥ 1 s
Interface
• Analog output (Not applicable to PROFIBUS PA option)
Signal range 4 to 20 mA
Upper Limit 20 to 22.5 mA adjustable
Fail signal 3.6 mA; 22 mA; 24 mA or last value
Load Max. 600 Ω; for HART®
1
communication min. 230 Ω
• Digital output
Function Configurable as a device status or limit value
(level, volume, mass)