A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 93
mmmmm
B: Troubleshooting
Operation Troubleshooting
Operating symptoms, probable causes, and resolutions.
S: 52 Fail-safe is activated. Possible
causes: 1) hardware failure; 2)
memory failure; 3) Fail-safe LOE
timer expired – possible causes:
faulty installation, antenna material
buildup, foaming/other adverse pro-
cess conditions, invalid calibration
range.
For 3): Correct configuration; ensure
installation is correct; no antenna
material buildup; adjust process
conditions to minimize foaming/
other adverse conditions; correct
calibration range. If fault persists, or
for 1) and 2), contact your local Sie-
mens representative.
S: 53 * Configuration lost: one or more
parameter settings have been lost.
This may occur after a firmware
upgrade causes user parameters to
be reset.
Restore user parameters using
SIMATIC PDM.
Symptom Cause Action
Display shows
S: 0 LOE
level or target is out of
range
check specifications
check
2.2.1.6. Low Calibration Pt.
(LOW CALIB. PT.)
increase
2.2.4.5.1. Echo Confidence
Display shows
S: 0 LOE
material build-up on
antenna
clean the antenna
re-locate SITRANS LR260
Display shows
S: 0 LOE
location or aiming:
poor installation
flange not level
check to ensure nozzle is vertical
•use
2.2.5.2. Auto False Echo
Suppression
and check
2.2.5.3. Auto
Suppression Range
to ensure nozzle
protrudes from end of vessel nozzle.
Display shows
S: 0 LOE
antenna malfunction:
temperature too high
physical damage
check
3.14.1. Current Internal
Temperature (CURR. INTERN.
TEMP.)
•relocate
General Fault Codes (Continued)
Code /
Icon
Meaning Corrective Action