User's Manual
Table Of Contents
- Table of Contents
- SITRANS LR260 Overview
- Specifications
- Installation
- Wiring
- Quick Start via local operation
- Operating via SIMATIC PDM
- Parameter Reference
- Pull-down menus via SIMATIC PDM
- Quick Start Wizard
- 1. Quick Start
- 2. Setup
- 3. Diagnostics
- 3.1. Echo Profile
- 3.14. Measured Values (MEAS. VALUES)
- 3.15. Remaining Device Lifetime (REMAIN. DEV. LIFE.)
- 3.15.2. Remaining Device Lifetime (REMAIN. LIFETIME)
- 3.15.3. Maintenance Required Limit (MAINT. REQ. LIMIT)
- 3.15.4. Maintenance Demanded Limit (MAINT. DEM. LIMIT)
- 3.15.5. Maintenance Alert Activation (ALERT ACTIVATION)
- 3.15.6. Total Expected Device Life (TOTAL EXP. LIFE)
- 3.15.7. Maintenance Status (MAINT. STAT.)
- 3.15.8. Acknowledge Status (ACK STATUS)
- 3.15.9. Acknowledge (ACK)
- 3.16. Remaining Sensor Lifetime (REMAIN SENS. LIFE.)
- 3.16.2. Remaining Sensor Lifetime (REMAIN. LIFETIME)
- 3.16.3. Maintenance Required Limit (MAINT. REQ. LIMIT)
- 3.16.4. Maintenance Demanded Limit (MAINT. DEM. LIMIT)
- 3.16.5. Maintenance Alert Activation (ALERT ACTIVATION)
- 3.16.6. Total Expected Sensor Life (TOTAL. EXP. LIFE)
- 3.16.7. Maintenance Status (MAINT. STATUS)
- 3.16.8. Acknowledge Status (ACK. STATUS)
- 3.16.9. Acknowledge (ACK.)
- 4. Service
- 4.1. Device Reset
- 4.2. Manufacture Date (MANUF. DATE)
- 4.3. LCD Fast Mode
- 4.4. LCD Contrast
- 4.11. Memory Test (MEM. TEST)
- 4.16.1. Time Last Serviced (TIME LAST SERV.)
- 4.16.2. Remaining Lifetime (REMAIN LIFETIME)
- 4.16.3. Maintenance Required Limit (MAINT. REQ. LIMIT)
- 4.16.4. Maintenance Demanded Limit
- 4.16.5. Maintenance Alert Activation
- 4.16.6. Total Service Interval
- 4.16.7. Maintenance Status (MAINT. STAT)
- 4.16.8. Acknowledge Status (ACK. STATUS)
- 4.16.9. Acknowledge (ACK.)
- 4.17. Calibration Interval (CALIB. INTERVAL)
- 4.17.1. Time Last Calibrated (TIME LAST CAL.)
- 4.17.2. Remaining Lifetime (REMAIN LIFETIME)
- 4.17.3. Maintenance Required Limit (MAINT. REQ. LIMIT)
- 4.17.4. Maintenance Demanded Limit (MAINT. DEM. LIMIT)
- 4.17.5. Maintenance Alert Activation
- 4.17.6. Total Calibration Interval (TOTAL CALIB. INTRV.)
- 4.17.7. Maintenance Status (MAINT. STAT.)
- 4.17.8. Acknowledge Status (ACK. STATUS)
- 4.17.9. Acknowledge (ACK.)
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Appendix E: Application Example
- Appendix F: HART Communications
- Appendix G: Firmware Revision History
- Glossary
- Index
- LCD menu structure
A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 93
mmmmm
B: Troubleshooting
Operation Troubleshooting
Operating symptoms, probable causes, and resolutions.
S: 52 Fail-safe is activated. Possible
causes: 1) hardware failure; 2)
memory failure; 3) Fail-safe LOE
timer expired – possible causes:
faulty installation, antenna material
buildup, foaming/other adverse pro-
cess conditions, invalid calibration
range.
For 3): Correct configuration; ensure
installation is correct; no antenna
material buildup; adjust process
conditions to minimize foaming/
other adverse conditions; correct
calibration range. If fault persists, or
for 1) and 2), contact your local Sie-
mens representative.
S: 53 * Configuration lost: one or more
parameter settings have been lost.
This may occur after a firmware
upgrade causes user parameters to
be reset.
Restore user parameters using
SIMATIC PDM.
Symptom Cause Action
Display shows
S: 0 LOE
level or target is out of
range
• check specifications
• check
2.2.1.6. Low Calibration Pt.
(LOW CALIB. PT.)
• increase
2.2.4.5.1. Echo Confidence
Display shows
S: 0 LOE
material build-up on
antenna
• clean the antenna
• re-locate SITRANS LR260
Display shows
S: 0 LOE
location or aiming:
• poor installation
• flange not level
• check to ensure nozzle is vertical
•use
2.2.5.2. Auto False Echo
Suppression
and check
2.2.5.3. Auto
Suppression Range
to ensure nozzle
protrudes from end of vessel nozzle.
Display shows
S: 0 LOE
antenna malfunction:
• temperature too high
• physical damage
• check
3.14.1. Current Internal
Temperature (CURR. INTERN.
TEMP.)
•relocate
General Fault Codes (Continued)
Code /
Icon
Meaning Corrective Action