Parameters
4.16.7. Maintenance Status (MAINT. STAT)
Read only. Displays the status of the Maintenance limits.
Open the menu View – Display Status, click on the Maintenance tab and check
the Service Schedule Status window.
4.16.8. Acknowledge Status (ACK. STATUS)
Read only. Acknowledges the current limit status.
Open the menu View – Display Status, click on the Maintenance tab and check
the Service Schedule Status window.
4.16.9. Acknowledge (ACK.)
Acknowledges the current limit status.
To acknowledge an alert via SIMATIC PDM:
a) Open the menu View – Device Status and click on the Maintenance tab.
b) In the Service Schedule Status section click on Acknowledge Warnings.
To acknowledge an alert via the handheld programmer:
a) Press RIGHT arrow twice to open parameter view and activate Edit Mode.
b) Press RIGHT arrow to acknowledge the alert.
4.17. Calibration Interval (CALIB. INTERVAL)
1)
The device tracks calibration intervals based on operating hours and monitors the
predicted lifetime to the next calibration. You can modify the Total Calibration Interval,
set schedules for Maintenance Alerts, and acknowledge them.
To access these parameters via SIMATIC PDM:
Open the menu Device – Maintenance and select the Calibration Interval tab.
After modifying values/units as required, click on Write to accept the change,
and Read to view the effect of the change.
Notes:
Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and
set up Maintenance/Service schedules, based on operating hours instead of a
calendar-based schedule. See also Remaining Device Lifetime, Service Interval,
and Calibration Interval.
The device operates in years. To view Remaining Sensor Lifetime parameters in
hours or days (via SIMATIC PDM only) see Total Expected Sensor Life.
1)
In device, Maintenance parameters are displayed in years. In SIMATIC PDM, mainte-
nance parameter units are user selectable.