A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 67
mmmmm
Parameters
3. Diagnostics
3.1. Echo Profile
Allows you to request the current echo profile either via the handheld programmer,
or via SIMATIC PDM. (For more detail see Echo Processing on page 96.)
To request a profile via SIMATIC PDM:
a. Open the menu View – Echo Profile.
b. If desired, press Save.
c. To view the saved profile, open the menu View – Show Echo Profile.
To request a profile via the handheld programmer:
a. In PROGRAM mode, navigate to LEVEL METER > DIAGNOSTICS > ECHO
PROFILE (3.1)
b. Press RIGHT arrow to request a profile.
c. In the Profile screen, press UP arrow to select the Transmit icon, and
Right ARROW to update the profile.
d. Press DOWN arrow to select the Exit icon, then Right ARROW to return to
previous menu.
3.14. Measured Values (MEAS. VALUES)
3.14.1. Current Internal Temperature (CURR. INTERN. TEMP.)
Read Only. Displays (in degrees C) the current temperature on the circuit board
recorded by the internal electronics.
3.14.2. Maximum Internal Temperature (MAX. INTERN. TEMP.)
Read Only. Displays (in degrees C) the maximum temperature recorded by the
internal electronics. The high and low values are maintained over a power
cycle.
transmit icon,
selected
exit icon,
deselected
exit icon, selected
transmit icon, deselected
Distance value
Algorithm selection (First echo)
TVT curve
material
level
Echo confidence
value