Page 64 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
mmmmm
Parameters
2.3.1.3. 4 mA Setpoint
Sets the process level corresponding to the 4 mA value. 4 mA always
defaults to 0, and mA Output Function determines whether this is a Level,
Space, or Distance measurement. Level and Space are measured as a
percentage of the difference between High Calibration Point and Low
Calibration Point. Distance is measured as a percentage of the Low
Calibration Point.
Enter the reading that is to correspond to a 4 mA output. Units are defined
in Sensor units for Level, Space, or Distance.
1)
Notes:
The mA Output Function is set independently from the setting in
Operation. Set Operation first and then mA Output Function.
(Operation resets mA Output Function to the same value.)
The mA Output Function controls the primary value and the loop
current for the SITRANS LR260. Use caution when changing the mA
Output function when connected to a HART network.
mA Output Function must be set to Manual before you can modify the
mA Output Value (2.3.1.1). Remember to restore your previous setting
after using mA Output Function.
Values
Range: 0 to 30 m
1)
Default: 0 m (set to value corresponding to 0% as defined
by mA Output Function)
Related
Parameters
mA Output Function
1)
4 mA setpoint can be set less than 0 m, and 20 mA setpoint can be set greater than
30 m when using handheld programmer. Recommended range is 0 to 30 m.
Level Space
Distance
Distance
Space
Level
20 mA
100%
0%
4 mA
High
Cal.
Point
4 mA
0%
100%
20 mA
4 mA
0%
100%
20 mA
Sensor
reference
point
Low
Calibration
Point