Page 50 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
mmmmm
Parameters
Quick Start Wizard
The Quick Start wizard groups together all the settings you need to configure for a simple
application. You can access it either via SIMATIC PDM, or via the handheld programmer.
Do not use the Quick Start wizard to modify individual parameters (perform customization
for your application only after the Quick Start has been completed).
1. Quick Start
1.1. Application Type
1.2. Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Use a setting just faster than the maximum filling or emptying rate (whichever is
greater). Slower settings provide higher accuracy; faster settings allow for more level
fluctuation.
1.3. Units
Select the units for the Quick Start variables (high and low calibration point, and level,
distance, or space).
1.4. Operation
Options
STEEL
Silo construction
CONCRETE
Default: STEEL
Options
SLOW 0.1 m/minute
MED 1.0 m/minute
FAST 10.0m/minute
Default: FAST
Options
M, CM, MM, FT, IN
Default: M
Options
NO SERVICE
Stops updating measurements and associated loop
current. Last valid measurement value is displayed.
LEVEL
Distance to material surface referenced from Low
Calibration Point (process empty level).
SPACE
Distance to material surface referenced from High
Calibration Point (process full level).
DISTANCE
Distance to material surface referenced from Sensor
Reference Point.
Default: DISTANCE