SIMATIC PDM
To set Service/Calibration schedules:
1) Open the menu Device – Maintenance, and click on the Service/Calibration
Schedule tab.
2) Modify desired values, and click on Write.
3) Click on Read, to see the effects of your modification.
4) If desired, activate alerts.
5) Click on Service Performed to reset the schedule.
Device Reset
Factory Defaults
Use Device Reset to reset all user parameters excluding device addresses, the parameter
block Identification, and Maintenance Schedule parameters to the default settings.
1. Open the menu Device – Device Reset and click on OK to perform a reset to Factory
Defaults.
2. After the reset is complete, upload parameters to the PC/PG. (If you are performing a
reset after replacing the device with a different instrument, do not upload
parameters to the PC/PG).
Configuration Flag Reset
To reset the configuration flag to zero, open the menu Device – Configuration Flag Reset
and execute a reset.
D/A (Digital/Analog) Trim
Allows you to trim the 4 mA and 20 mA points in order to calibrate the mA output.
Open the menu Device – D/A Trim. You will be prompted to attach a calibrated meter and
enter the values at 4 mA and at 20 mA.
Select AnalogOut
Allows you to select one of three operation types:
•level
•space
•distance
1) Open the menu Device – Select AnalogOut.
2) The Select AnalogOut window displays the current setting: click on OK.
3) Select the desired option, and click on OK.
4) The Select AnalogOut window displays the new setting: click on OK.