A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 15
mmmmm
Installation
Nozzle location (continued)
Provide easy access for viewing the display and programming via the hand
programmer.
Provide an environment suitable to the housing rating and materials of construction.
Provide a sunshield if the instrument will be mounted in direct sunlight.
Orientation in a vessel with obstructions
Polarization reference point
ambient temperature
–40 °C to +80 °C (–40 °F to +176 °F)
process temperature (with FKM O-ring)
non-pressure version: –40 to +200 °C (–40 to +392 °F)
pressure rated version: –40 to +80 °C (–40 to +176 °F)
polarization
reference
point
polarization
axis
display
For best results on a vessel with
obstructions, orient the front or back of
the device toward the obstructions.