User's Manual

Table Of Contents
A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 7
mmmmm
Specifications
Influence of ambient temperature < 0.003%/K (average over full temperature range,
referenced to maximum range)
Dielectric constant of material measured
Minimum dK = 1.6 (depending on antenna and application type)
Memory:
non-volatile EEPROM
no battery required.
Interface
Analog output
signal range 4 to 20 mA (±0.02 mA accuracy)
upper limit 20 to 23 mA adjustable
fail signal 3.6 mA to 23 mA; or last value
load Max. 600
Communication: HART
Load 230 to 600 230 to 500 when connecting a
coupling module
Max. Line Length multi-wire: 1500 m (4921 ft)
•Protocol HART
1)
, Version 5.1
Configuration Siemens SIMATIC PDM (PC), or
Siemens Milltronics infrared handheld programmer,
or HART handheld communicator
Display (local)
2)
graphic LCD, with bar graph representing level
Mechanical (Easy Aimer model)
Materials
flange and horn 304 stainless steel
Process Connections:
universal flanges
3)
2"/50 mm, 3"/80 mm, 4"/100 mm, 6"/150 mm
Horn:
•2" horn
•3" horn
•4" horn
Enclosure
construction aluminum, polyester powder-coated
conduit entry 2 x M20x1.5, or 2 x ½" NPT
ingress protection Type 4X/NEMA 4X, Type 6/NEMA 6, IP67, IP68
Dust cap (optional)
2" PTFE, pipe clamp connection, O.D. 50 mm (1.97")
3" PTFE, pipe clamp connection, O.D. 75 mm (2.95")
4" PTFE, pipe clamp connection, O.D. 100 mm (3.94")
1)
HART® is a registered trademark of HART Communication Foundation.
2)
Display quality will be degraded in temperatures below –25 °C (–13 °F) and above +65 °C (+149 °F).
3)
Universal flange mates with EN 1092-1 (PN16)/ASME B16.5 (150 lb)/JIS 2220 (10K) bolt hole
pattern.