B: Troubleshooting
Reading does not
change, but the level
does
SITRANS LR260 process-
ing wrong echo, i.e. vessel
wall, or structural member
re-locate SITRANS LR260
ensure nozzle protrudes 10 mm min.
rotate instrument 90°. See
Orientation in a vessel with
obstructions
on page 15.
•use
2.2.5.2. Auto False Echo
Suppression
and
2.2.5.3. Auto
Suppression Range
Measurement is con-
sistently off by a con-
stant amount
setting for
2.2.1.6. Low
Calibration Pt. (LOW
CALIB. PT.)
not correct
setting for
2.2.1.25.
Sensor Offset
not correct
check distance from sensor
reference point to
2.2.1.6. Low
Calibration Pt. (LOW CALIB. PT.)
check
2.2.1.25. Sensor Offset
Screen blank
power error check nameplate rating against
voltage supply
check power wiring or source
too much load resistance remove something from the loop, or
increase supply voltage
Reading erratic
echo confidence weak refer to
2.2.4.5.1. Echo Confidence
•use
2.2.5.2. Auto False Echo
Suppression
and
2.2.5.3. Auto
Suppression Range
use foam deflector or stillpipe
material filling re-locate SITRANS LR260
Reading response
slow
2.2.7.2. Fill Rate (FILL RATE/
MIN)
setting incorrect
increase measurement response if
possible
Reads correctly but
occasionally reads
high when vessel is
not full
detecting close range
echo
build up near top of
vessel or nozzle
nozzle problem
clean the antenna
•use
2.2.5.2. Auto False Echo
Suppression
and
2.2.5.3. Auto
Suppression Range
Level reading lower
than material level
material is within Near
Range zone
multiple echoes
processed
decrease
2.2.1.12. Near Range
(minimum value depends on
antenna type)
raise SITRANS LR260
•ensure
2.2.4.1.1. Algorithm
is set to F
(First echo)