Radar Transmitters SITRANS LR260 (HART) Operating Instructions 12/2013 SITRANS
Safety Guidelines: Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the connected equipment. These warning notices are accompanied by a clarification of the level of caution to be observed. Qualified Personnel: This device/system may only be set up and operated in conjunction with this manual. Qualified personnel are only authorized to install and operate this equipment in accordance with established safety practices and standards.
Table of Contents SITRANS LR260 Overview ................................................................................................5 Specifications ....................................................................................................................6 Power............................................................................................................................................. 6 Performance...............................................................................
Table of Contents mmmmm Quick Start Wizard via the handheld programmer .....................................................................32 Requesting an Echo Profile ..............................................................................................................34 Device Address ...................................................................................................................................34 Auto False Echo Suppression ..................................................
Appendix C: Maintenance .............................................................................................95 Unit Repair and Excluded Liability ..................................................................................................95 Appendix E: Application Example ..............................................................................107 Flour in steel storage vessel, level measurement ....................................................................
Table of Contents mmmmm Notes iv
Safety Notes Special attention must be paid to warnings and notes highlighted from the rest of the text by grey boxes.1) WARNING: relates to a caution symbol on the product, and means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. WARNING1: means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage.
CE Electromagnetic Compatibility (EMC) Conformity SITRANS LR260 mmmmm This equipment has been tested and found to comply with the following EMC Standards: EMC Standard Title CISPR 11:2004/EN 55011:1998+A1:1999&A2:2002, CLASS B Limits and methods of measurements of radio disturbance characteristics of industrial, scientific, and medical (ISM) radio-frequency equipment.
The Manual Notes: • Please follow the installation and operating procedures for a quick, trouble-free installation and to ensure the maximum accuracy and reliability of your SITRANS LR260. • This manual applies to the SITRANS LR260 (HART) only. This manual will help you set up your SITRANS LR260 for optimum performance. We always welcome suggestions and comments about manual content, design, and accessibility. Please direct your comments to techpubs.smpi@siemens.com.
SITRANS LR260 mmmmm Abbreviations and Identifications Short form Long Form A/D Analog to digital CE / FM / CSA Conformité Européenne / Factory Mutual / Canadian Standards Association Ci Internal capacitance D/A Digital to analog DAC Digital Analog Converter DCS Distributed Control System dK dielectric constant ESD Electrostatic Discharge HART Highway Addressable Remote Transducer Ii Input current mA Io Output current mA IS Intrinsically Safe Li Internal inductance mH F Desc
SITRANS LR260 Overview SITRANS LR260 is a 2-wire 25 GHz pulse radar level transmitter for continuous monitoring of solids and liquids in storage vessels including extreme levels of dust and high temperatures, to a range of 30 m (98.4 ft). The instrument consists of an electronic component coupled to a horn antenna with an integral Easy Aimer and flange or threaded connection for quick and easy positioning. A dust cover or air purging are available as options.
Specifications Note: Siemens Milltronics makes every attempt to ensure the accuracy of these specifications but reserves the right to change them at any time. Power Nominal 24 V DC with max.
Influence of ambient temperature < 0.003%/K (average over full temperature range, referenced to maximum range) Dielectric constant of material measured • Minimum dK = 1.6 (depending on antenna and application type) Memory: • non-volatile EEPROM • no battery required. Interface Analog output • signal range • fail signal • load Communication: HART • Load • Max. Line Length • Protocol 4 to 20 mA (±0.02 mA accuracy) upper limit 20 to 23 mA adjustable 3.6 mA to 23 mA; or last value Max.
Air Purge Connection • equipped with female 1/8" NPT fitting Weight • standard model < 8.14 kg (17.9 lb) including 4" flange and standard Easy Aimer with 4" horn antenna Mechanical (Threaded Connection model) • threaded connection materials 2" NPT (ASME B1.20.1), R (BSPT, EN 10226-1) or G (BSPP, EN ISO 228-1) 316L /1.4404 or 316L /1.4435 stainless steel PTFE emitter Environmental Notes: Specifications mmmmm • Check Approvals on page 8, for the specific configuration you are about to use or install.
FM/CSA Class I, Div. 1, Groups A, B, C, D, Class II, Div. 1, Groups E, F, G, Class III T4 FM/CSA Class I, Div 2, Groups A, B, C, D T5 SABS ARP0108 Ex ia IIC T4 Ga INMETRO: DNV 12.
Dimensions: Notes: • Process temperature and pressure capabilities are dependent upon information on the process device tag. • Signal amplitude increases with horn diameter, so use the largest practical size. enclosure/electronics nameplate Specifications mmmmm 268 mm (10.6") process device tag universal slotted flange sensor reference point optional extension 2" horn: 163 mm (6.4") 3" horn: 228 mm (9.0") horn 4" horn: 285 mm (11.2") 49 mm (1.9") 74.5 mm (2.9") 97.5 mm (3.
Threaded Horn Antenna with extension Notes: • Process temperature and pressure capabilities are dependent upon information on the process connection tag. • Signal amplitude increases with horn diameter, so use the largest practical size. • Optional extensions can be installed below the threads. 50 mm (2.0") 1/2" NPT cable entry (or alternatively, M20 cable gland) threaded cover 167 mm (6.6") 154 mm (6.1") retaining collar 196 mm (7.7") 159 mm (6.26") optional 100 mm (4.0 ") horn extension 28 mm (1.
Threaded Horn dimensions Antenna Type Antenna O.D. Height to sensor reference pointa) 1-1/2" threaded connection 2" threaded connection Beam Angleb) Measurement Range 3" threaded connection 2" Horn 47.8mm (1.88") N/A 166mm (6.55") 180mm (7.09") 15 degrees 20m (65.6 ft) 3" Horn 74.8mm (2.94") N/A 199mm (7.85") 213mm (8.39") 10 degrees 20m (65.6 ft) 4" Horn 94.8mm (3.73") N/A 254mm (10") 268mm (10.55") 8 degrees a) 20m (65.
Installation WARNINGS: • Installation shall only be performed by qualified personnel and in accordance with local governing regulations. • SITRANS LR260 is to be used only in the manner outlined in this manual, otherwise protection provided by the device may be impaired. • Never attempt to loosen, remove, or disassemble process connection or instrument housing while vessel contents are under pressure.
Mounting location Notes: • Correct location is key to a successful application. • Avoid reflective interference from vessel walls and obstructions by following the guidelines below. Nozzle design • The end of the horn must protrude a minimum of 10 mm (0.4”) to avoid false echoes being reflected from the nozzle. • Optional antenna extensions: 100 mm (3.93"), 200 mm (70.9"), 500 mm (19.69"), 1000 mm (39.4")1) Nozzle location Min. clearance: 10 mm (0.4") min.
Nozzle location (continued) • • • Provide easy access for viewing the display and programming via the hand programmer. Provide an environment suitable to the housing rating and materials of construction. Provide a sunshield if the instrument will be mounted in direct sunlight.
Easy Aimer Note: When the Easy Aimer ball is loosened, the device is free to tilt to a maximum of 30°. Easy Aimer ball locking bolts top clamping plate (upper socket) 0.38" (10 mm) bottom clamping plate (lower socket) customer gasket as required [recommended thickness 1.5 to 1.8 mm (0.06 to 0.07")] customer mounting plate, as required ø 4" (102 mm) min., central opening 285 mm (11.22") Installation mmmmm 30° max. 1.
Air Purging System (Optional) For more frequent cleaning, a purging system can be installed between the flange and the horn antenna. The system provides an 1/8" inlet (female thread) on the flange where cooling air or cleaning fluid passes through the flange and exits the inside of the horn to clean it. The customer will supply the purging medium by a manual or automatic valve system. This option is only available with the universal flange for purging shown on page 18.
Universal Slotted Flange WARNING: The user is responsible for the selection of bolting and gasket materials which will fall within the limits of the flange and its intended use and which are suitable for the service conditions. number of slotted bolt holes bolt hole circle max. diameter bolt hole circle min. diameter flange O.D. thickness section A-A Installation mmmmm Slotted Flange Dimensions1) Pipe Size Flange O.D. 2" or 50 mm 3" or 80 mm 4" or 100 mm 6" or 150 mm 6.50" (165 mm) 7.87" (200 mm) 9.
Dust Cap (Optional) Note: The dust cap must be removed before using the Air Purge feature (see Air Purging System (Optional) on page 17). The dust cap fits onto the end of the horn and prevents the buildup of dust and other process material inside the horn. • It is particularly useful for applications in areas of high humidity, or with bulk solids with a high moisture content. Three sizes are available, to fit the standard 2", 3" and 4" horns. • Installation 1. Thoroughly clean inside the horn.
Wiring Power WARNINGS: The DC input terminals shall be supplied from a source providing electrical isolation between the input and output, in order to meet the applicable safety requirements of IEC 61010-1. All field wiring must have insulation suitable for rated voltages. Connecting SITRANS LR260 WARNINGS: • Check the nameplate on your instrument, to verify the approval rating. • Use appropriate conduit seals to maintain IP or NEMA rating.
1. 2. 3. 4. Strip the cable jacket for approximately 70 mm (2.75") from the end of the cable, and thread the wires through the gland1). Connect the wires to the terminals as shown: the polarity is identified on the terminal block. Ground the instrument according to local regulations. Tighten the gland to form a good seal.
1. Intrinsically Safe wiring Device nameplate (ATEX/IECEX/INMETRO/C-TICK) SITRANS LR260 7MLxxxx-xxxxx-xxxx- BOX 1 Serial No: GYZ / S1034567 Encl.: NEMA / TYPE 4X, 6, IP67, IP68 Amb.Temp.: – 40°C to 80°C Power Rating: 24 V Nom., 30 V HART Max., 4 - 20 mA Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts KCC-REM-S49 SITRANSLR II 1 G D, 1/2 D , 2 D Ex ia IIC T4 Ga Ex ta IIIC T100°C Da SIRA 11ATEX2348X IECEx SIR 11.
Passive Shunt Diode Barriers Note: A well regulated supply voltage is required. Manufacturer Part Number MTL 787SP+ (Dual Channel) MTL 7787P+ (Dual Channel) Stahl 9001/01-280-100-10 (Single Channel) Stahl 9002/01-280-110-10 (Dual Channel) How to select a passive barrier for SITRANS LR260 To make sure that the barrier safety description is suitable for the LR260 Intrinsically Safe (IS) input parameters, carry out the following calculations: Re-e Rloop = = max.
2. Non-Sparking wiring SITRANS LR260 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Encl.: NEMA / TYPE 4X, 6, IP67, IP68 Amb. Temp.: – 40°C to 80°C Power Rating: 24 V Nom., 30 V 4 – 20 mA II 3 G Ex nA IIC T4 Gc SIRA 09ATEX4155X KCC-REM-S49 SITRANSLR Max.
Instructions specific to hazardous area installations (Reference European ATEX Directive 94/9/EC, Annex II, 1/0/6) The following instructions apply to equipment covered by certificate number SIRA 11ATEX2348X, and SIRA 09ATEX4155X. 1) 2) For use and assembly, refer to the main instructions. The equipment is certified for use as Category 1GD, 1/2D, 2D equipment per SIRA 11ATEX2348X; and Category 3G equipment per SIRA 09ATEX4155X.
Quick Start: local mmmmm Quick Start via local operation SITRANS LR260 carries out its level measurement tasks according to settings made via parameters. The settings can be modified locally via the Local User Interface (LUI) which consists of an LCD display and a handheld programmer. A Quick Start Wizard provides an easy step-by-step guide to help you configure the device for a simple application.
PROGRAM mode display current item number current item pointer Parameter view parameter name parameter number parameter value/ selection Edit view A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 27 mmmmm • A visible menu bar indicates the menu list is too long to display all items. • A band halfway down the current menu menu bar indicates the current item is halfway down the list.
Quick Start: local mmmmm Handheld Programmer (Part No. 7ML1930-2AJ) The programmer is ordered separately. C Key functions in Measurement mode Key Function Updates the loop current. Result New value is displayed in LCD secondary region. Updates internal enclosure temperature New value is displayed in LCD secondary region. reading. Page 28 Updates echo confidence value. New value is displayed in LCD secondary region. Updates distance measurement. New value is displayed in LCD secondary region.
Programming SITRANS LR260 Change parameter settings and set operating conditions to suit your specific application. • See Operating via SIMATIC PDM on page 36 for remote operation. Programming via the handheld programmer Parameter menus Note: For the complete list of parameters with instructions, see Parameter Reference on page 49. Parameters are identified by name and organized into function groups, then arranged in a 5-level menu structure (seeLCD menu structure on page 121).
Quick Start: local mmmmm 2. Navigating: key functions in Navigation mode Notes: • In Navigation mode, ARROW keys move to the next menu item in the direction of the arrow. • For Quick Access to parameters via the handheld programmer, press Home , and then enter the menu number, for example: 2.2.1. Key Name UP or DOWN arrow Menu level Function menu or parameter Scroll to previous or next menu or parameter. menu Go to first parameter in the selected menu, or open next menu.
Changing a numeric value parameter name b. Press RIGHT arrow to open parameter view. The current value is displayed. c. Press RIGHT arrow again to open Edit mode. The current value is highlighted. d. Key in a new value. e. Press RIGHT arrow to accept it. The LCD returns to parameter view and displays the new selection. parameter number current value Key functions in Edit mode Key Name Function Selecting options Scrolls to item.
Quick Start: local mmmmm Quick Start Wizard via the handheld programmer Notes: • The Quick Start Wizard settings are inter-related and changes apply only after you click on Yes in Apply?at the end of the Quick Start steps. • Do not use the Quick Start wizard to modify individual parameters: see instead Parameter Reference on page 49 (perform customization for your application only after the Quick Start has been completed). 1. Quick Start a.
1.4. Operation space level low calibration point 1) Operation types NO SERVICE SITRANS LR260 stops updating measurements and associated loop current. Last valid measurement is displayed. LEVEL Distance to material surface referenced from Low Calibration Point (process empty level). SPACE Distance to material surface referenced from High Calibration Point (process full level). DISTANCE Distance to material surface referenced from Sensor Reference Point. Default: DISTANCE 1.5.
Quick Start: local mmmmm Requesting an Echo Profile a. In PROGRAM mode, navigate to: LEVEL METER > DIAGNOSTICS > ECHO PROFILE (3.1) b. Press RIGHT arrow profile. c. In the Profile screen, press UP arrow to request a transmit icon, deselected to select the Transmit icon, and Right ARROW update the profile. d. DIAGNOSTICS 3.1 ECHO PROFILE MEAS. VALUES REMAIN. DEV. LIFE REMAIN. SENS. LIFE to Press DOWN arrow to select the Exit icon, then Right ARROW to return to previous menu.
Level application example SITRANS LR260 Sensor Reference Point 0.5 m High Calibration Point 16 m level Low Calibration Point Quick Start Setting APPLICATION Description STEEL RESPONSE RATE SLOW UNITS m Response rate = 0.1 m/minute. OPERATION LEVEL Material level referenced from Low Calibration Point. LOW CALIBRATION POINT 16 Process empty level. HIGH CALIBRATION POINT 0.5 Process full level. APPLY? (CHANGES) YES Save new settings. Fill rate = 0.09 m/minute (Low Cal Pt.
Operating via SIMATIC PDM Note: For a complete list of parameters with instructions, see Parameter Reference starting on page 49. SIMATIC PDM mmmmm SIMATIC PDM is a software package used to commission and maintain SITRANS LR260 and other process devices. Please consult the operating instructions or online help for details on using SIMATIC PDM. (You can find more information at www.fielddevices.com: go to Products and Solutions > Products and Systems > Communications and Software > Process Device Manager.
Electronic Device Description (EDD) Note: SITRANS LR260 requires the EDD for SIMATIC PDM version 6.0 with SP3 or higher. You can locate the EDD in Device Catalog, under Sensors/Level/Echo/Siemens Milltronics/ SITRANS LR260. Check the product page of our website at: www.siemens.com/LR260, under Downloads, to make sure you have the latest version of SIMATIC PDM, the most recent Service Pack (SP) and the most recent hot fix (HF). If you need to install a new EDD, see Configuring a new device below.
Quick Start Notes: • The Quick Start wizard settings are inter-related and changes apply only after you click on Transfer at the end of step 4. • Do not use the Quick Start Wizard to modify individual parameters: see instead Parameter Reference on page 49. (Perform customization only after the Quick Start has been completed.) • Initial Quick Start parameter values are not default values and do not necessarily reflect the current device configuration.
Step 2 – Application Type Select the application type and the operation, then click on NEXT. Set the parameters, and click on NEXT.
Step 4 – Summary SIMATIC PDM mmmmm Check parameter settings, and click on BACK to return and revise values, or TRANSFER to transfer values to the device. The message Quick Start was successful will appear. Click on OK, then click on OK again to synchronize with the device.
Changing parameter settings using SIMATIC PDM Notes: • For a complete list of parameters, see Parameter Reference on page 49. • Clicking on Cancel during an upload from device to SIMATIC PDM will result in some parameters being updated. 1. 2. 3. Parameters accessed via pull-down menus You have access to a number of functions via pull-down menus from the menu bar under Device or View. For a complete list see Pull-down menus via SIMATIC PDM on page 49.
Display SIMATIC PDM mmmmm To compare outputs in real time, open the menu View – Display. Echo profile Notes: • Double click on each axis and record the Xscale and Data Scale values, so that you can restore the default view by resetting to these values. • You can save a profile or delete a saved profile. • After saving a profile, open menu View – Echo profile.
Trend Notes: • Double click on each axis and record the Xscale and Data Scale values, so that you can restore the default view by resetting to these values.
Auto False Echo Suppression Notes: • If possible, adjust Auto False Echo Suppression parameters with an empty or almost empty vessel. • Set Auto False Echo Suppression and Auto Suppression Range during startup, if possible. • Before adjusting these parameters, rotate the instrument for best signal (lower false-echo amplitude). SIMATIC PDM mmmmm SITRANS LR260 first learns the echo profile. Then the learned profile, or part of it, is used to screen out false echoes.
TVT Shaper Note: Double click on each axis and record the Xscale and Data Scale values, so that you can restore the default view by resetting to these values. This feature allows you to manually adjust the TVT curve to avoid false echoes caused by obstructions. (For an explanation, see Auto False Echo Suppression on page 98.) Open the menu Device – TVT Shaper TVT curve echo profile • • • • Press Measure to refresh the echo profile and load the current TVT curve from the device.
Device Status Open the menu View – Device Status to view Device Status, Maintenance, and Diagnostics status. • See Diagnostics on page 67. • See Remaining Device Lifetime (REMAIN. DEV. LIFE.) on page 68 and Remaining Sensor Lifetime (REMAIN SENS. LIFE.) on page 71.
To set Service/Calibration schedules: 1) 2) 3) 4) 5) Open the menu Device – Maintenance, and click on the Service/Calibration Schedule tab. Modify desired values, and click on Write. Click on Read, to see the effects of your modification. If desired, activate alerts. Click on Service Performed to reset the schedule. Device Reset Factory Defaults 2. Open the menu Device – Device Reset and click on OK to perform a reset to Factory Defaults. After the reset is complete, upload parameters to the PC/PG.
Simulate AO (Analog Output) Allows you to input a simulated value in order to test the functioning of the mA connections during commissioning or maintenance of the device. To simulate a user-defined mA value: SIMATIC PDM mmmmm 1. Open the menu Device – Simulate AO. 2. Select Other, enter the simulated value, and click on OK. The message ’Field Device fixed at new value’ appears. Click on OK. 3.
Parameter Reference Notes: • See Programming via the handheld programmer on page 29 for detailed instructions. • Do not use the handheld programmer at the same time as SIMATIC PDM, or erratic operation may result. • Mode toggles between PROGRAM and Measurement Modes. • For Quick Access to parameters via the handheld programmer, press Home , then enter the menu number, for example: 2.2.1. • In Navigation mode, ARROWs navigate the menu in the direction of the arrow.
Quick Start Wizard The Quick Start wizard groups together all the settings you need to configure for a simple application. You can access it either via SIMATIC PDM, or via the handheld programmer. Do not use the Quick Start wizard to modify individual parameters (perform customization for your application only after the Quick Start has been completed). 1. Quick Start 1.1. Application Type Options STEEL Silo construction CONCRETE Default: STEEL 1.2.
Operation types sensor reference point (flange face) high calibration point distance space level low calibration point 1.5. Low Calibration Point (LOW CALIB. PT.) Distance from Sensor Reference to Low Calibration Point: usually process empty level. (See Operation types for an illustration.) Range: 0.0000 to 30.000 m Default: horn type dependent Values Range: 0.0000 to 30.000 m Default: 0.0 m Related parameters 2.2.1.12.
2. Setup Notes: • See Programming via the handheld programmer on page 29 or Operating via SIMATIC PDM on page 36 for instructions. • Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated. • Values shown in the following tables can be entered via the handheld programmer. 2.1. Device 2.1.1. Firmware Revision (FIRMWARE REV.) Corresponds to the software or firmware that is embedded in the SITRANS LR260. Read only. 2.1.2. Loader Revision (LOADER REV.
Operation types sensor reference point (flange face) high calibration point distance2 space level low calibration point 2.2.1.6. Low Calibration Pt. (LOW CALIB. PT.) Distance from Sensor Reference to Low Calibration Point (corresponding to Low Level Point). Units are defined in Sensor Units.1) 2) SITRANS LR260 with Easy Aimer sensor reference point (flange face) high calibration point low calibration point Values Range: 0 to 30 m. Default dependent on horn type 2.2.1.7. High Calibration Pt.
2.2.1.8. Limit Sensor Value When On, limits the sensor value to the range between Low Calibration Point and High Calibration Point. ON, OFF. Default: On Options 2.2.1.12. Near Range The range in front of the device (measured from the sensor reference point) within which any echoes will be ignored, defined in sensor units. This is sometimes referred to as blanking or a dead zone. The factory setting is 50 mm past the end of the horn and the range is dependent on the horn type and any extensions.
2.2.4. Echo Processing (ECHO PROC.) 2.2.4.1. Echo select 2.2.4.1.1. Algorithm Selects the algorithm to be applied to the echo profile to extract the true echo. Options * ALF A L F AL AF LF BLF BL BF L TF Area Largest First Echo Area Largest Echo First echo Area Largest Area First Largest First Best of Largest or First echo Best Largest Best First Last echo True First 2.2.4.1.2. Echo Position Defines where on the echo the distance measurement is determined.
2.2.4.2. Sampling Provides a method of checking the reliability of a new echo before accepting it as the valid reading, based on numbers of samples above or below the currently selected echo. 2.2.4.2.1. Echo Lock Selects the measurement verification process. Options Lock Off * Maximum Verification Material Agitator Total Lock (not recommended for radar) Related parameters 2.2.7.2. Fill Rate (FILL RATE/MIN) 2.2.7.3. Empty rate (EMPTY RATE/MIN) 2.2.4.2.4. Up Sampling 2.2.4.2.5. Down Sampling 2.2.4.2.6.
2.2.4.2.6. Window A "distance window" centered on the echo1), used to derive the reading. When a new measurement is in the window, the window is re-centered and the reading is calculated. Values Range: 0 to 30 m Default: 0 When the value is 0, the window is automatically calculated after each measurement. • For slower Measurement Response values, the window is narrow. • For faster Measurement Response values, the window becomes progressively wider.
2.2.4.3.7. Reform Echo Smooths jagged peaks in the echo profile. Reforms fragmented echoes into one echo. 0 = OFF Values Related parameters greater = wider Range: 0 to 50 mS Recommended: 5 to 20 mS; higher is not recommended. 2.2.4.1.1. Algorithm 2.2.4.3.6. Narrow Echo Filter 2.2.4.1.5. Echo Marker 2.2.4.5. Noise 2.2.4.5.1. Echo Confidence Indicates echo reliability. It displays the echo confidence of the last measurement.
2.2.5. TVT (Auto False Echo Suppression) setup (TVT SETUP) First SITRANS LR260 learns the echo profile. Then the learned profile, or part of the learned profile, is used to screen out false echoes. See Before and After Auto False Echo Suppression on page 60 for examples. 2.2.5.1.
Before Auto False Echo Suppression Level (db) default TVT echo marker learned TVT curve Level (db) Parameters mmmmm After Auto False Echo Suppression level false echo echo marker 2.2.5.3. Auto Suppression Range Defines the endpoint of the Learned TVT distance. Units are defined in sensor units. Values a. b. c. Range: 0.00 to 30.00 m Default: 1.00 m Press RIGHT ARROW to open Edit mode. Enter the new value and press RIGHT ARROW to accept it. Set 2.2.5.2. Auto False Echo Suppression . 2.2.5.4.
2.2.5.6. TVT Type Selects the TVT shaping type to be applied to the echo profile to extract the true echo. Options Short Curved Short Flat Long Flat Long Smooth Front Long Smooth Slopes * Long Smooth 2 Short Curved 2 2.2.6. TVT shaper A breakpoint on the TVT curve. This allows you to reshape the TVT curve to avoid unwanted echoes. There are 40 breakpoints arranged in five groups. To adjust the TVT curve after shape definition, see Shaper Mode on page 60. 2.2.6.1.
2.2.7. Rate 2.2.7.1. Response Rate Note: Changing Response Rate resets Fill Rate, Empty Rate, Damping Filter and Echo Lock. Sets the reaction speed of the device to measurement changes. Related Response parameRate ters Options * Fill Rate Empty Rate Damping Filter Echo Lock slow 0.1 m/min. 0.1 m/min. 10 s max. verifications medium 1 m/min. 1 m/min. 10 s max. verifications fast 10 m/min. 10 m/min. 0s max.
Enter a value slightly greater than the vessel’s maximum emptying rate, in Sensor Units per minute. Options Meters/Minute Slow 0.1 Medium 1 Fast 10 2.2.8. Transducer Block (TB) Values For diagnostic purposes 2.2.8.2. Level Measurement (LEVEL MEAS.) The value for level. Read Only. In SIMATIC PDM, open the menu View – Display, and select the tab Display. 2.2.8.3. Space Measurement (SPACE MEAS.) The value for space. Read Only.
Notes: • The mA Output Function is set independently from the setting in Operation. Set Operation first and then mA Output Function. (Operation resets mA Output Function to the same value.) • The mA Output Function controls the primary value and the loop current for the SITRANS LR260. Use caution when changing the mA Output function when connected to a HART network. • mA Output Function must be set to Manual before you can modify the mA Output Value (2.3.1.1).
2.3.1.4. 20 mA Setpoint Sets the process level corresponding to the 20 mA value. 20 mA always defaults to 100%, and mA Output Function determines whether this is a Level, Space, or Distance measurement. Level and Space are measured as a percentage of the difference between High Calibration Point and Low Calibration Point: Distance is measured as a percentage of Low Calibration Point.
2.3.1.8. 20 mA Output Trim Calibrates the 20 mA output. The mA output of the device is pre-calibrated; however, 20 mA Output Trim can be used to trim remote displays or inputs. Values Related parameters Range: 18.0 to 24.0 (mA) 4 mA Output Trim Steps: 1. Set mA Output Function to Manual. 2. Set mA Output Value to 20 mA. 3. Attach a calibrated meter and check the output at the terminals; record the remote reading in mA. 4. Enter this value in 20 mA Output Trim. 5.
3. Diagnostics 3.1. Echo Profile Allows you to request the current echo profile either via the handheld programmer, or via SIMATIC PDM. (For more detail see Echo Processing on page 96.) To request a profile via SIMATIC PDM: a. Open the menu View – Echo Profile. b. If desired, press Save. c. To view the saved profile, open the menu View – Show Echo Profile.
3.14.3. Minimum Internal Temperature (MIN. INTERN. TEMP.) Read Only. Displays (in degrees C) the minimum temperature recorded by the internal electronics. The high and low values are maintained over a power cycle. 3.15. Remaining Device Lifetime (REMAIN. DEV. LIFE.)1) Notes: • Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up Maintenance/Service schedules, based on operating hours instead of a calendar-based schedule.
3.15.1. Total Device Operating Time (TOTAL OP. TIME) Read only. The amount of time the device has been operating in selected units. 3.15.2. Remaining Device Lifetime (REMAIN. LIFETIME) Read only. Remaining Lifetimein selected units. 3.15.3. Maintenance Required Limit (MAINT. REQ. LIMIT) If the Remaining Lifetimeis equal to or less than this limit, the device sets Maintenance Required. Values Range: 0 to 20 years Default: 0.
3.15.6. Total Expected Device Life (TOTAL EXP. LIFE) Note: The device always operates in years. Changing the units affects only the parameter view of the Remaining Device Lifetime parameters in SIMATIC PDM. Allows you to override the factory default Values a) Unitsa): hours, days, years Range: 0 to 20 years Default: 10 years Units are selectable only via SIMATIC PDM. 3.15.7. Maintenance Status (MAINT. STAT.) Read only. Displays the status of the Maintenance limits.
3.16. Remaining Sensor Lifetime (REMAIN SENS. LIFE.)1) Notes: • Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up Maintenance/Service schedules, based on operating hours instead of a calendar-based schedule. See also Remaining Device Lifetime, Service Interval, and Calibration Interval. • The device operates in years. To view Remaining Sensor Lifetime parameters in hours or days (via SIMATIC PDM only) see Total Expected Sensor Life.
3.16.2. Remaining Sensor Lifetime (REMAIN. LIFETIME) Read only. Total Expected Sensor Life less Total Sensor Operating Time in selected units. 3.16.3. Maintenance Required Limit (MAINT. REQ. LIMIT) If Remaining Sensor Lifetime is equal to or less than this limit, the device sets Maintenance Required. Values Range: 0 to 20 years Default: 0.164 years To modify the value via PDM, open the menu Device – Maintenance and click on Remaining Sensor Lifetime. a) Modify limit values as required. b) Enable 3.16.
3.16.6. Total Expected Sensor Life (TOTAL. EXP. LIFE) Allows you to override the factory default. Unitsa): hours, days, years Range: 0 to 20 years Default: 10 years Values a) Units are selectable only via SIMATIC PDM. 3.16.7. Maintenance Status (MAINT. STATUS) Read only. Displays the status of the Maintenance limits Via SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab and check the Sensor Lifetime Status window. 3.16.8. Acknowledge Status (ACK. STATUS) Read only.
4.3. LCD Fast Mode Note: Affects Measurement mode only: has no effect on Navigation mode. Enables a faster rate of measurement from the device by disabling most of the display area. Only the bar graph will be refreshed when LCD Fast Mode is set to ON. When Fast Mode is set to ON, there is a delay of 30 minutes with no pressed buttons before actions result. Values ON or OFF Default: OFF 4.4. LCD Contrast The factory setting is for optimum visibility at room temperature and in average light conditions.
4.16. Service Interval1) Notes: • Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up Maintenance/Service schedules, based on operating hours instead of a calendar-based schedule. See also Remaining Device Lifetime, Service Interval, and Calibration Interval. • The device operates in years. To view Remaining Sensor Lifetime parameters in hours or days (via SIMATIC PDM only) see Total Expected Sensor Life.
4.16.2. Remaining Lifetime (REMAIN LIFETIME) Read only. Total Service Interval less Time Last Serviced in selected units. 4.16.3. Maintenance Required Limit (MAINT. REQ. LIMIT) If Remaining Lifetime is equal to or less than this limit, the device sets Maintenance Required. Values Range: 0 to 20 years Default: 0.164 years a) Modify limit values as required. b) Set Maintenance Alert Activation to the desired option. 4.16.4.
4.16.7. Maintenance Status (MAINT. STAT) Read only. Displays the status of the Maintenance limits. Open the menu View – Display Status, click on the Maintenance tab and check the Service Schedule Status window. 4.16.8. Acknowledge Status (ACK. STATUS) Read only. Acknowledges the current limit status. Open the menu View – Display Status, click on the Maintenance tab and check the Service Schedule Status window. 4.16.9. Acknowledge (ACK.) Acknowledges the current limit status.
• Click on Calibration Performed to restart the timer and clear any fault messages. 4.17.1. Time Last Calibrated (TIME LAST CAL.) Time elapsed since last calibration Can be reset to zero via the handheld programmer (after performing a service). • Parameters mmmmm To reset to zero: • In SIMATIC PDM, open the menu Device – Maintenance, click on the Calibration Interval tab, and click on Calibration Performed to restart the timer and clear any fault messages.
4.17.5. Maintenance Alert Activation Note: To modify this parameter via SIMATIC PDM it must be accessed via the pull-down menu Device – Maintenance. Allows you to enable a maintenance limit. Values * Timer off On - no limits On - limit 1 On - limits 1 and 2 On - limit 2 a) First set the limit values in Maintenance Required Limit/Maintenance Demanded Limit. b) Select the desired Alert Activation option. 4.17.6. Total Calibration Interval (TOTAL CALIB. INTRV.) Note: The device always operates in years.
5. Communication 5.1. Device Address Sets the device address or poll ID on a HART network. Any address other than 0 will cause the output current to be a fixed value, and the current will not indicate the reading. Values Range: 0 to 15 Default: 0 To set the address via the handheld programmer: • Press RIGHT arrow to activate PROGRAM mode and open menu level 1. • Press DOWN arrow item 5). repeatedly to navigate to COMMUNICATION (menu • Press RIGHT arrow to display the parameter list.
6. Security 6.1. Lock Note: • Do not lose this number value. • This lock affects only the handheld programmer. A remote master can change configuration if Communication Control is set to allow this. Prevents any changes to parameters via the handheld programmer. Handheld programmer Values Related parameter Range: 1 to 9999 * 1954 (unlock value) Any other value Off (enables local programming) On (disables local programming) Unlock value (6.2.
mmmmm Parameters Page 82 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
page 77 for those parameters. Parameter Name (Parameter Number) 20 mA Output Trim 20 mA Setpoint (2.3.1.8.) 66 (2.3.1.4.) 4 mA Output Trim 4 mA Setpoint Page Number 65 (2.3.1.7.) 65 (2.3.1.3.) 64 Algorithm (2.2.4.1.1.) Application Type (1.1.) Auto False Echo Suppression (2.2.5.2.) Auto Suppression Range (2.2.5.3.) Communication Control (5.2.) Current Internal Temperature (CURR. INTERN. TEMP.) 55 50 59 60 80 67 (3.14.1.) Damping Filter (2.2.4.3.2.) Device Address (5.1.
A: Parameter List mmmmm Parameter Name (Parameter Number) High Calibration Pt. (HIGH CALIB. PT.) (2.2.1.7.) Page Number 53 Language (7.) LCD Contrast (4.4.) 81 74 LCD Fast Mode 74 (4.3.) Level Measurement (LEVEL MEAS.) Limit Sensor Value (2.2.1.8.) (2.2.8.2.) 63 54 Loader Revision (LOADER REV.) (2.1.2.) Lock (6.1.) Low Calibration Pt. (LOW CALIB. PT.) (2.2.1.6.) mA Output Function (mA OUTPUT FUNC.) (2.3.1.2.) mA Output Value (2.3.1.1.) Manufacture Date (MANUF. DATE) (4.2.
Parameter Name (Parameter Number) (6.2.) A5E32337683 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS 56 57 Page 85 mmmmm Up Sampling (2.2.4.2.4.) Window (2.2.4.2.6.
mmmmm A: Parameter List Page 86 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
Appendix B: Troubleshooting Communication Troubleshooting Generally: 2. 3. 4. . Verify that the wiring connections are correct. Check that the device address is correctly set. If you continue to experience problems, go to our website at: www.siemens.com/processautomation, and check the FAQs for SITRANS LR260, or contact your local Siemens Milltronics representative. Specifically: 1. 2. 3.
Device Status Icons B: Troubleshooting mmmmm Icon Page 88 Priority Level Meaning 1 • Maintenance alarm • Measurement values are not valid 2 • Maintenance warning: maintenance demanded immediately • Measured signal still valid 3 • Maintenance required • Measured signal still valid 1 • Process value has reached an alarm limit 2 • Process value has reached a warning limit 3 • Process value has reached a tolerance limit 1 • Configuration error • Device will not work because one or more param
Priority Level Icon Meaning (Continued) • Data exchanged • No data exchange • Write access enabled Notes: • If more than one fault is present, the device status indicator and text for each fault alternate at 2 second intervals. • Some faults cause the device to go to Failsafe mode (Fault 52). These are indicated with an asterisk (*). General Fault Codes Code / Icon Meaning Corrective Action S: 0 * The device was unable to get a measurement within the Fail-safe LOE Timer period.
General Fault Codes (Continued) B: Troubleshooting mmmmm Code / Icon Meaning Corrective Action S: 3 Device is nearing its lifetime limit according to the value set in Maintenance Required Limit. Replacement is recommended. S: 4 Device is nearing its lifetime limit according to the value set in Maintenance Demanded Limit. Replacement is recommended. S: 6 Sensor is nearing its lifetime limit according to the value set in Maintenance Required Limit. Replacement is recommended.
General Fault Codes (Continued) Code / Icon Meaning Corrective Action Calibration interval as defined in Maintenance Required Limit has expired. Perform calibration. S: 18 Calibration interval as defined in Maintenance Demanded Limit has expired. Perform calibration. * Internal device failure caused by a RAM memory error. Repair required: contact your local Siemens representative. S: 29 * EEPROM damaged. Repair required: contact your local Siemens representative. S: 31 * Flash error.
General Fault Codes (Continued) B: Troubleshooting mmmmm Code / Icon Meaning Corrective Action S: 36 * Unable to start microwave module. Cycle power. If fault persists, contact your local Siemens representative. S: 37 * Measurement hardware problem. Cycle power. If fault persists, contact your local Siemens representative. S: 38 * Microwave module hardware failure: unable to calculate distance measurement. Cycle power.
General Fault Codes (Continued) Code / Icon S: 52 * Fail-safe is activated. Possible causes: 1) hardware failure; 2) memory failure; 3) Fail-safe LOE timer expired – possible causes: faulty installation, antenna material buildup, foaming/other adverse process conditions, invalid calibration range. For 3): Correct configuration; ensure installation is correct; no antenna material buildup; adjust process conditions to minimize foaming/ other adverse conditions; correct calibration range.
B: Troubleshooting Symptom mmmmm Cause Action (Continued) Reading does not change, but the level does SITRANS LR260 processing wrong echo, i.e. vessel wall, or structural member • re-locate SITRANS LR260 • ensure nozzle protrudes 10 mm min. • rotate instrument 90°. See Measurement is consistently off by a constant amount • setting for 2.2.1.6.
Appendix C: Maintenance SITRANS LR260 requires no maintenance or cleaning under normal operating conditions. Under severe operating conditions, the horn antenna may require periodic cleaning. If cleaning becomes necessary: • Note the antenna material and the process medium, and select a cleaning solution that will not react adversely with either. • Remove the instrument from service and wipe the antenna clean using a cloth and suitable cleaning solution.
Appendix D: Technical Reference Principles of Operation SITRANS LR260 is a 2-wire 25 GHz pulse radar level transmitter for continuous monitoring of solids and liquids1). Radar level measurement uses the time of flight principle to determine distance to a material surface. The device transmits a signal and waits for the return echo. The transit time is directly proportional to the distance from the material. Pulse radar uses polarized electromagnetic waves.
Higher order mathematical techniques and algorithms are used to provide intelligent processing of microwave reflection profiles. This “knowledge based” technique produces superior performance and reliability. Time Varying Threshold (TVT) Curves A Time Varying Threshold (TVT) curve hovers above the echo profile to screen out unwanted reflections (false echoes). sensor reference point High Cal. Pt. =0 obstruction at 1.2 m TVT curve material level at 2.9 m false echo true echo Low Cal. Pt.
The options are Center, CLEF (Constrained Leading Edge Fit), Hybrid or Rising Edge. CLEF uses the leading edge of the echo. It can be used to compensate for materials with a low dK value, which may cause the vessel bottom to be reported as the level instead of the actual material level, in low level conditions. CLEF range is the level below which the CLEF algorithm will be used: above this level the Center algorithm is used.
Far Range In applications where the base of the vessel is conical or parabolic, a reliable echo may be available below the vessel empty distance, due to an indirect reflection path. Increasing the range extension to 30% or 40% can provide stable empty vessel readings.1) Measurement Response The measurement response (response rate) limits the maximum rate at which the display and output respond to changes in the measurement.
When the instrument is returned to Measurement mode, the transceiver resumes operation. The reading and mA output default to the last measurement taken. The reading and associated outputs migrate to the current process level at a rate controlled by the response rate (2.2.7.1). If SITRANS LR260 is left in PROGRAM mode for 10 minutes without input, it automatically reverts to Measurement mode. Damping A damping filter smooths out the response to a sudden change in level.
Fail-safe Timer Fail-safe Timer determines the time to elapse after the last valid reading before a Failsafe state is activated. Fail-safe Material Level determines the level to be reported when the Fail-safe timer expires. Upon receiving a reliable echo, the loss of echo condition is aborted, the Service Required icon and error message are cleared, and the reading and mA output return to the current level. Fail-safe Material Level When the 2.4.1.
Process Pressure/Temperature derating curves Notes: • The Process Device Tag shall remain with the process pressure boundary assembly1. In the event the instrument package is replaced, the Process Device Tag shall be transferred to the replacement unit. • SITRANS LR260 units are hydrostatically tested, meeting or exceeding the requirements of the ASME Boiler and Pressure Vessel Code and the European Pressure Equipment Directive.
Horn Antenna 2" (50 mm), 3" (80 mm), 4" (100 mm), and 6" (150 mm) Flanged Versions Process Connection Series: • 51262 or 51263 series flange. • Ensure your instrument has the process identification Tag showing one of this series, and 25589 stamped on flange. WARNING: Never attempt to loosen, remove or disassemble process connection or instrument housing while vessel contents are under pressure.
Loop power Typical Connection Drawing Note: Loop voltage is the voltage at the terminals of the power supply (not the voltage at the terminals of the device). loop resistance RL loop voltage power supply LR260 + loop current IL Allowable operating area of SITRANS LR260 Loop Voltage versus Loop Resistance Power Supply Requirements 1000 (30.0v,845 ohms) Loop Resistance - RL D: Technical Reference mmmmm 900 800 RL=44.6*VL - 493.4 700 600 500 400 (17.
Startup Behavior Notes: • SITRANS LR260 is designed to start reliably with a power supply capable of delivering at least 25 mA. • When connected to a power supply with a current limit of < 25 mA, the LR260 may not start reliably.
mmmmm D: Technical Reference Page 106 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
Appendix E: Application Example Note: In the applications illustrated below, values are for example purposes only. You can use this example as setup references. Enter the values in the parameter tables to select the corresponding functions. Configure the basic settings using the Quick Start wizard parameters. (These parameters are inter-related, and changes take effect only after you select YES in step 7 to apply changes.
Parameter type Parameter No. and Name 1.1. Application Type 1.2. Response Rate 1.3. Units 1.4. Operation 1.5. Low Calibration Point Quick Start (LOW CALIB. PT.) Wizard 1.6. High Calibration Point parameters (HIGH CALIB. PT.) 1.7. Apply? (Apply changes) 2.4.1. Fail-safe Timer Independent parameters 2.4.2. Fail-safe Material Level (FAILSAFE MAT. LE) Function STEEL SLOW Slow =0.1 m/minute M meters LEVEL Level 16 16 m 0.5 0.5 m YES Transfers Quick Start settings to device.
SITRANS LR260 can be configured over the HART network using either the HART Communicator 375 by Fisher-Rosemount, or a software package. The recommended software package is the SIMATIC Process Device Manager (PDM) by Siemens. SIMATIC PDM This software package is designed to permit easy configuration, monitoring, and troubleshooting of HART devices. The HART EDD for SITRANS LR260 was written with SIMATIC PDM in mind and has been extensively tested with this software.
F: HART Communications mmmmm HART Communicator 375 Menu Structure Note: HART Communicator 375 is supported by SITRANS LR260 HART. The menu structure is aligned with the menu structure of SIMATIC PDM. SITRANS LR260 Identification Tag Description Message Installation Date Device order number Setup Device Input Page 110 Firmware Revision Loader Revision Hardware Revision Sensor Calibration Antenna Sensor Units Operation Low Calibration Pt. High Calibration Pt.
Output Fail-Safe mA Output Shaper 1-9 (Shaper Points 1-9) Shaper 10-18 (Shaper Points 10-18) Shaper 19-27 (Shaper Points 19-27) Shaper 28-36 (Shaper Points 28-36) Shaper 37-40 (Shaper Points 37-40) Response Rate Fill Rate per minute Empty Rate per minute mA output function 4 mA setpoint 20 mA setpoint Minimum mA limit Maximum mA limit Failsafe timer Failsafe Material level Failsafe Level Diagnostics Measured Values Current Internal Temp Max Internal Temp Min Internal Temp Remaining Device Lifetime Tota
F: HART Communications mmmmm Calibration Interval Time Elapsed Since Last Calibration Remaining Lifetime Maintenance Required Limit Maintenance Demanded Limit Maintenance Alert Activation Calibration Acknowledge Total Calibration Interval Communication Communication control Security Language Page 112 Lock Unlock value Language SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
SITRANS LR260 conforms to HART rev.
Firmware Rev. EDD Rev. 1.00.01 1.00.04 Date Dec.
Glossary accuracy: degree of conformity of a measure to a standard or a true value. algorithm: a prescribed set of well-defined rules or processes for the solution of a problem in a finite number of steps. ambient temperature: the temperature of the surrounding air that comes in contact with the enclosure of the device. antenna: an aerial which sends out and receives a signal in a specific direction. There are four basic types of antenna in radar level measurement, horn, parabolic, rod, and waveguide.
derating: to decrease a rating suitable for normal conditions according to guidelines specified for different conditions. dielectric: a nonconductor of direct electric current.1) dielectric constant (dK): the ability of a dielectric to store electrical potential energy under the influence of an electric field. Also known as Relative Permittivity. An increase in the dielectric constant is directly proportional to an increase in signal amplitude.
multiple echoes: secondary echoes that appear as double, triple, or quadruple echoes in the distance from the target echo. Near Blanking: see Blanking nozzle: a length of pipe mounted onto a vessel that supports the flange. parameters: in programming, variables that are given constant values for specific purposes or processes. polarization: the property of a radiated electromagnetic wave describing the time-varying direction and amplitude of the electric field vector.
stilling-well: see stillpipe. two wire radar: a low-energy radar. Can be loop powered, analog, intrinsically safe 4 to 20 mA, or a digital (BUS) transmitter. TVT (time varying threshold): a time-varying curve that determines the threshold level above which echoes are determined to be valid. Glossary mmmmm waveguide antenna: a hollow, metallic tube that transmits a microwave signal to the product target.
Index A Abbreviations and Identifications list 4 activating LR260 26 analog output fail signal 7 in Measure mode 99 load 7 signal range 7 approvals 8 Auto False Echo Suppression explanation 98 setup 59 via PDM 43 B beam angle values 12 blanking (see Near Range) 98 C cables requirements 20 cleaning instructions 95 communication load 7 max.
maintenance 46 cleaning 95 maintenance settings 67 Measurement Response explanation 99 mounting handheld programmer access 15 housing construction 15 nozzle design 14 nozzle location 14 on vessel with obstructions 15 sunshield recommended 15 N Near Range explanation 98 O operating principles 96 P parameter settings adjust via PDM 36 performance specifications 6 polarization reference point 15 power 6 power source requirements 20 power supply requirements loop voltage vs.
Appendix C: menu chart LCD menu structure SITRANS LR260 (HART) – OPERATING INSTRUCTIONS Page 121 SITRANS LR260 (HART) – OPERATING INSTRUCTIONS A5E32337683
Appendix C: menu chart LCD menu structure Notes: • In Navigation mode ARROW keys navigate the menu in the direction of the arrow.. • See See Parameter Reference on page 43. for detailed information and instructions. 1. QUICK START 1.1 APPLICATION TYPE 1.2 RESPONSE RATE 1.3 UNITS 1.4 OPERATION 1.5 LOW CALIB. PT. 1.6 HIGH CALIB. PT. 1.7 APPLY? 2. SETUP 2.1 DEVICE 2.1.1 FIRMWARE REV 2.1.2 LOADER REV 2.1.3 HARDWARE REV 2.2 INPUT 2.2.1 SENSOR CALIB. 2.2.1.1 ANTENNA 2.2.1.4 SENSOR UNITS 2.2.1.5 OPERATION 2.2.
3. 4. 5. A5E32337683 SITRANS LR260– OPERATING INSTRUCTIONS Page 123 mmmmm 7. LCD menus 6. 3.15.3 MAINT REQ LIMIT 3.15.4 MAINT DEM LIMIT 3.15.5 ALERT ACTIVATION 3.15.6 TOTAL EXP. LIFE 3.15.7 MAINT STAT 3.15.8 ACK STATUS 3.15.9 ACK 3.16 REMAIN. SENS. LIFE 3.16.1 SENSOR OP-TIME 3.16.2 REMAIN. LIFETIME 3.16.3 MAINT REQ LIMIT 3.16.4 MAINT DEM LIMIT 3.16.5 ALERT ACTIVATION 3.16.6 TOTAL EXP. LIFE 3.16.7 MAINT STAT 3.16.8 ACK STATUS 3.16.9 ACK SERVICE 4.1 DEVICE RESET 4.2 MANUF. DATE 4.3 LCD FAST MODE 4.
mmmmm LCD menus Page 124 SITRANS LR260 – OPERATING INSTRUCTIONS A5E32337683
IQ300IX.
www.siemens.com/processautomation For more information www.siemens.com/level www.siemens.com/continuous-weighing Siemens AG Industry Sector 1954 Technology Drive P.O. Box 4225 Peterborough, ON Canada K9J 7B1 Subject to change without prior notice A5E32337683 Rev. AA © Siemens AG 2013 email: techpubs.smpi@siemens.com www.siemens.