3.7.3. Event Diagnosis (EVENT DIAG.)
Allows you to assign one of the diagnostic options listed below to any of the
events listed in Event Index(3.7.1.). This allows you to tailor a device response
appropriate for your particular process.(Event Diagnosis affects Condensed
Acyclic Diagnostics and Cyclic Extended Diagnostics
1)
).
To assign a diagnosis to a particular event via the handheld programmer:
a) Go to Condensed Status Mode (COND. STAT. MODE)(3.6.1.) and select Yes
to enable Condensed Mode.
b) Go to Event Index(3.7.1.) and enter the event index number corresponding
to a particular event.
c) Go to Event Status (EVENT. STAT)(3.7.2.) and choose a Status option.
d) Go to Event Diagnosis (EVENT DIAG.)(3.7.3.) and choose a Diagnosis option
from the table above.
To assign a status to a particular event via SIMATIC PDM:
a) Go to Level Meter > Diagnostics > Condensed Status Setup, and select
Yes to enable Condensed Status Mode.
b) Go to Level Meter > Diagnostics > Condensed Status.
c) Select the Diagnosis line for the selected Event, then choose a Diagnosis
option from the associated pull-down menu.
Event Diagnosis Options
Status OK
Maintenance Required
Maintenance Demanded
* Maintenance alarm
Invalid process condition
Function check or simulation
1)
See
Condensed Mode Diagnosis
on page 126 for more detail.