Page 72 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS A5E32337685
mmmmm
Parameters
3. Diagnostics
Use these parameters to set up schedules for calibration and maintenance. The device
will track itself based on operating hours instead of a calendar-based schedule, and will
monitor its predicted lifetime.
The maintenance warnings and alarms are communicated via either the Status or
Condensed Status bytes. This information can be integrated into any Asset Management
system. We recommend that you use SIMATIC PCS7 Asset Management Software in
conjunction with SIMATIC PDM.
3.1. Fault Reset
Clears the following fault (known as Acknowledge Fault in PDM.)
To clear a fault using the handheld programmer:
Enter the fault code number then press RIGHT Arrow.
To clear a fault via SIMATIC PDM:
Open the menu View – Acknowledge Faults.
Select the fault to be cleared from the pull-down menu in Extended
Diagnostics.
Click on Transfer to clear the fault.
3.2. Echo Profile
You can request the current echo profile either via the handheld programmer (see
next page) or via SIMATIC PDM. (For more details see Echo Processing on page
103.)
To request a profile via SIMATIC PDM:
a. Open the menu View – Echo Profile.
b. If desired, press Save.
c. To view the saved profile, open the menu View – Show Echo Profile.
Fault Code Description
S12 Internal Temperature High
Fault Code Description
S3 Device Lifetime Limit 1 (Maintenance Required)
S4 Device Lifetime Limit 2 (Maintenance Demanded)
S6 Sensor Lifetime Limit 1 (Maintenance Required)
S7 Sensor Lifetime Limit 2 (Maintenance Demanded)
S8 Device Service Limit 1 (Maintenance Required)
S9 Device Service Limit 2 (Maintenance Demanded)
S17 Calibration Schedule Limit 1 (Maintenance Required
S18 Calibration Schedule Limit 2 (Maintenance Demanded)
S25 Internal Error (Maintenance Demanded)