A5E32337685 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 59
mmmmm
Parameters
2.2.3.2.4. Window
A “distance window” centered on the echo
1)
is used to derive the
reading. When a new measurement is in the window, the window is
re-centered and the reading is calculated.
When the value is 0, the window is automatically calculated after
each measurement.
For slower Measurement Response values, the window is
narrow.
For faster Measurement Response values, the window
becomes progressively wider.
2.2.3.3. Filtering
2.2.3.3.1. Shots
The number of echo profile samples averaged to produce a
measurement.
2.2.3.3.2. Narrow Echo Filter (NARROW ECHO FIL.)
Filters out echoes of a specific width
To remove a false echo from the Echo Profile, take its width in mm
and multiply it by 0.013. Enter the result.
For example, to filter out a spike with 500 mm width, enter 6 or 7 (the
closest integer product of 500 x 0.013).
When a value is keyed in, the nearest acceptable value is entered.
1)
See
Echo Lock
on page 104 for more detail.
Values
Range: 0 to 31.5 m
Default: 0 m
Note: The Echo Lock window is stored as a standard sample, but
displayed in sensor units. Any value entered for the Echo Lock
window will be rounded to the nearest sample.
Values
Range: 1 to 25
Default: 10
Values
Range: 0 to 255
0 = OFF
greater = wider
Default: 0 (OFF)