Page 48 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS A5E32337685
mmmmm
SIMATIC PDM
Maintenance
You can set schedules for:
maintenance of the device based on its projected lifetime
maintenance of the sensor based on its projected lifetime
•service
calibration
To set Device/Sensor Maintenance schedules:
1. Open the menu Device – Maintenance, and click on the Remaining Device/
Sensor Lifetime tab.
2. If desired, activate alerts for either or both of Maintenance Required/
Maintenance Demanded.
3. Modify desired values, and click on Write.
4. Click on Read, to see the effects of your modification.
5. Click on Snooze to add a year to the Total Expected Device Life.
To set Service/Calibration schedules:
1. Open the menu Device – Maintenance, and click on the Service/Calibration
Interval tab.
2. If desired, activate alerts.
3. Modify desired values, and click on Write.
4. Click on Read, to see the effects of your modification.
5. Click on Service Performed to reset the schedule.