SIMATIC PDM
Simulate Input
Allows you to simulate the sensor value which is input to the Level Transducer Block. This
tests everything between the Level Transducer Block and Output.
1. Open the menu Device – Simulation, and select Simulation (Input).
2. To enable simulation select Fixed or Ramp.
If you select Ramp, enter the Ramp start and end, step length and number of
steps. Click on Transfer.
If you select Fixed, enter the simulated value and click on Transfer.
3. After simulation is complete, disable simulation.
Device Reset
To perform a reset to Factory Defaults:
1. Open the menu Device – Device Reset, and click on Factory Defaults.
2. After the reset is complete, click on Close, then upload parameters to the PC/PG.
(If you are performing a reset after replacing the device with a different
instrument, do not upload parameters to the PC/PG.)
Write Locking
If Write Locking is enabled, the data can be viewed but not modified. See
Write Locking
(6.1.)
on page 88 for more details.
To enable/disable Write Locking:
1. Open the menu Device – Write Locking and select On or Off.
2. Click on Transfer.
Reset options Result
Factory Defaults Resets all parameters excluding Device Address , Tag,
Description, Message, and Installation Date to the
manufacturer’s default settings.
Standard Defaults Resets all parameters excluding Device Address, Description,
Message, and Installation Date to the PROFIBUS default
settings
Informational Resets parameter Tag
Functional Resets parameters that control device behavior, such as Low
Calibration Pt.
Warm Start Has the same effect as recycling power to the device
Reset Device Address
to 126
Resets the PROFIBUS device address to 126
If the address lock was on, will disable the lock.