A5E32337685 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 35
mmmmm
Quick Start: local
Requesting an Echo Profile
a) In PROGRAM mode, navigate to:
LEVEL METER > DIAGNOSTICS >
ECHO PROFILE (3.2)
b) Press RIGHT arrow to request a
profile.
c) In the Profile screen, press UP arrow
to select the Transmit icon, and
Right ARROW to update the profile.
d) Press DOWN arrow to select the Exit
icon, then Right ARROW to return to
previous menu.
Device Address
The unique address of the device on the network (also called PROFIBUS address).
a) In PROGRAM mode, navigate to: Level Meter > Communication > Device Address.
b) Press RIGHT arrow , RIGHT arrow , to open parameter view and enable Edit
mode.
c) If required, key in a new value and press RIGHT arrow to accept it. The LCD
displays the new value.
d) Press Mode to return to Measurement mode.
Auto False Echo Suppression
If SITRANS LR260 displays a false high level, or the reading is fluctuating between the
correct level and a false high level, you can use the Auto False Echo Suppression
parameters to prevent false echo detection. See TVT (Auto False Echo Suppression)
setup (2.2.4.) on page 61 for instructions.
Values
0 - 126. Default: 126