A5E32337685 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 33
mmmmm
Quick Start: local
Quick Start Wizard via the handheld programmer
1. Quick Start
a. Point the programmer at the display (from a maximum distance of 500 mm [1.64 ft]),
then press RIGHT arrow to activate PROGRAM mode and open menu level 1.
b. Press RIGHT arrow twice to navigate to menu item 1.1 and open parameter
view.
c. Press RIGHT arrow to open Edit mode or DOWN arrow to accept default
values and move directly to the next item.
d. To change a setting, scroll to the desired item or key in a new value.
e. After modifying a value, press RIGHT arrow to accept it and press DOWN arrow
to move to the next item.
1.1. Application Type
1.2. Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Use a setting just faster than the maximum filling or emptying rate (whichever is
greater).
1.3. Units
Select the units for the Quick Start variables (high and low calibration point, and level,
distance, or space)
Notes:
The Quick Start Wizard is a complete package and the settings are inter-related.
Do not use the Quick Start wizard to modify individual parameters: see instead
Parameter Reference
on page 51 (perform customization for your application after
the quick start has been completed).
Options
STEEL
Silo construction
CONCRETE
Default: STEEL
Options
SLOW 0.1 m/minute
MED 1.0 m/minute
FAST 10.0 m/minute
Default: FAST
Options
M, CM, MM, FT, IN
Default: M