Page 130 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS A5E32337685
mmmmm
G: Communications PA
Acyclic Data
SITRANS LR260 supports up to four simultaneous connections by a Class 2 Master
(C2 connection). It supports one connection by a Class 1 Master (C1 connection).
In order for a Class 1 Master to read parameters from an instrument, it needs to know the
slot and absolute index of the parameter.
The parameters are all listed in SIMATIC PDM under Help. If you do not have SIMATIC
PDM you can download the EDD and reference the HTML help file directly.
(continued on next page)
S: 32
The IDENT number used in
communications, and number
selected by the Ident Number
Selector do not correspond.
Ensure value of the Ident
number selector is correct for
the network configuration. If it is
correct, the device needs to be
re-parameterized by the PLC.
4
0
S:33
Factory calibration for the
internal temperature sensor
has been lost.
Repair required: contact your
local Siemens representative.
1
S:34
Factory calibration for the
device has been lost.
Repair required: contact your
local Siemens representative.
2
S:35
Factory calibration for the
device has been lost.
Repair required: contact your
local Siemens representative.
3
S:36
Unable to start microwave
module.
Cycle power. If error persists,
contact your local Siemens
representative.
4
S:37
Measurement hardware
problem.
Cycle power. If error persists,
contact your local Siemens
representative.
5
S:38
Microwave module hardware
failure: unable to calculate
distance measurement.
Cycle power. If fault persists,
contact your local Siemens
representative: repair required.
6
S:43
Factory calibration for the radar
receiver has been lost.
Repair required: contact your
local Siemens representative.
5
3
Acyclic Extended Diagnostics /General Fault Codes (cont’d)
LCD
display
Meaning (cont’d) Corrective Action
Byte
(cont’d)
Bit