User's Manual

Table Of Contents
A5E32337685 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 129
mmmmm
G: Communications PA
S:11
Internal temperature sensor
failure.
Return the device to the factory.
1
(cont’d)
3
S:12 *
Internal temperature of device
has exceeded specifications: it
is operating outside its
temperature range.
Lower the ambient temperature
enough to cool the device.
Fault code will persist until a
manual reset is performed using
PDM or the LCD interface.
4
S:14
Upper and lower input values
(Process Value Scale) for AIFB1
are the same.
Check configuration for AIFB1.
Ensure that Upper Value and
Lower Value (Process Value
Scale) are not the same.
6
S:15
Upper and lower input values
(Process Value Scale) for AIFB2
are the same.
Check configuration for AIFB2.
Ensure that Upper Value and
Lower Value (Process Value
Scale) are not the same.
7
S: 17
Calibration interval as defined
in Maintenance Required Limit
has expired.
Perform calibration.
2
1
S: 18
Calibration interval as defined
in Maintenance Demanded
Limit has expired.
Perform calibration. 2
S: 25
Internal device error
Reset power. If error persists,
contact your local Siemens
representative.
3
1
S:28
Internal device failure caused
by a RAM memory error.
Repair required: contact your
local Siemens representative.
4
S:29
EEPROM damaged.
Repair required: contact your
local Siemens representative.
5
S:31
Flash error.
Repair required: contact your
local Siemens representative.
7
Acyclic Extended Diagnostics /General Fault Codes (cont’d)
LCD
display
Meaning (cont’d) Corrective Action
Byte
(cont’d)
Bit