User's Manual

Table Of Contents
A5E32337685 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 7
mmmmm
Specifications
Mechanical (Threaded Connection model)
threaded connection 2" NPT (ASME B1.20.1), R (BSPT, EN 10226-1) or
G (BSPP, EN ISO 228-1)
materials 316L /1.4404 or 316L /1.4435 stainless steel
PTFE emitter
Environmental
location indoor/ outdoor
altitude 2000 m (6562 ft) max.
ambient temperature 40 to +80 °C (40 to +176 °F)
relative humidity suitable for outdoor
Type 4X/NEMA 4X, Type 6/NEMA 6, IP67 enclosure
(see note below)
installation category I
pollution degree 4
Process
•temperature
1)
-40 to +200 °C (-40 to +392 °F)
2)
(at process connection with FKM O-ring)
pressure (vessel)
1
Refer to
Process Pressure/Temperature derating curves
on
page 108.
Approvals
•General CSA
US/C
, FM, CE
Radio Europe (R&TTE), FCC, Industry Canada
Hazardous ATEX II 1GD, 1/2D, 2D Ex ia IIC T4 Ga, Ex ta IIIC T100 °C Da
ATEX II 3G Ex nA IIC T4 Gc
IECEx SIR 11.0153X Ex ia IIC T4 Ga, Ex ta IIIC T100 °C Da
FM/CSA Class I, Div. 1, Groups A, B, C, D, Class II, Div. 1,
Groups E, F, G, Class III T4
FM/CSA Class I, Div 2, Groups A, B, C, D T5
SABS ARP0108 Ex ia IIC T4 Ga
Notes:
Check
Approvals
on page 7, for the specific configuration you are about to use or
install.
Use appropriate conduit seals to maintain IP or NEMA rating.
1)
The maximum temperature is dependent on the ambient temperature and vessel
pressure. For more detail, or for other configurations, see
Process Pressure/
Temperature derating curves
on page 108.
2)
Pressure rated version for maximum process temperature of 80 °C (176 °F)
Note: The device nameplate lists the approvals that apply to your device.