User's Manual

Table Of Contents
Page 6 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS A5E32337685
mmmmm
Specifications
Dielectric constant of material measured
Minimum dK = 1.6 (depending on antenna and application type)
Memory
non-volatile EEPROM
no battery required.
Interface
PROFIBUS PA
configuration Siemens SIMATIC PDM (PC), or
Siemens Milltronics infrared hand-held programmer
(see
Programmer (infrared keypad)
on page 8)
display (local)
1)
graphic LCD, with bar graph (representing level)
Mechanical (Easy Aimer model)
Materials
flange and horn flange and horn: 304 stainless steel
Process Connections:
universal flanges
2)
2"/50 mm, 3"/80 mm, 4"/100 mm, 6"/150 mm
Horn:
•2" horn
•3" horn
•4" horn
Enclosure
construction aluminum, polyester powder-coated
conduit entry 2 x M20x1.5, or 2 x ½” NPT
ingress protection Type 4X/NEMA 4X, Type 6/NEMA 6, IP67, IP68 (see
note below)
Dust cap (optional)
2" PTFE, pipe clamp connection, O.D. 50 mm (1.97")
3" PTFE, pipe clamp connection, O.D. 75 mm (2.95")
4" PTFE, pipe clamp connection, O.D. 100 mm (3.94")
Air Purge Connection
equipped with female 1/8” NPT fitting
Weight
standard model < 8.14 kg (17.9 lb) including 4" flange and standard
Easy Aimer with 4" horn antenna
1)
Display quality will be degraded in temperatures below –25 °C (–13 °F) and
above +65 °C (+149 °F).
2)
Universal flange mates with EN 1092-1 (PN16)/ASME B16.5 (150 lb)/JIS 2220 (10K)
bolt hole pattern.