User's Manual

Table Of Contents
Page 106 SITRANS LR260 (PROFIBUS PA)– OPERATING INSTRUCTIONS A5E32337685
mmmmm
D: Technical Reference
Far Range
In applications where the base of the vessel is conical or parabolic, a reliable echo may
be available below the vessel empty distance, due to an indirect reflection path.
Increasing the range extension to 30% or 40% can provide stable empty vessel readings.
Measurement Response
The measurement response (Response Rate (1.2.)) limits the maximum rate at which the
display and output respond to changes in the measurement. Once the real process fill/
empty rate (m/s) is established, a response rate can be selected that is slightly higher
than the application rate. The response rate automatically adjusts the filters that affect
the output response rate.
There are three preset options: slow, medium, and fast.
Damping
A damping filter smooths out the response to a sudden change in level. This is an
exponential filter and the engineering unit is always in seconds. The setting can be
modified in Filter Time Constant (FILTER TIME CONST) (2.3.1.4.).
Loss of Echo (LOE)
A loss of echo (LOE) occurs when the calculated measurement is judged to be unreliable
because the echo confidence value has dropped below the echo confidence threshold.
Response
Rate (1.2.)
Fill Rate (FILL RATE/
MIN) (2.2.6.2.)/
Empty rate (EMPTY
RATE/MIN) (2.2.6.3.)
Filter Time
Constant (FILTER
TIME CONST)
(2.3.1.4.)
a)
a)
Numbering depends on AIFB selected.
Shots (2.2.3.3.1.)
Slow 0.1 m/min 60 s 25
Medium 1 m/min 10 s 10
Fast * 10 m/min 0 s 10
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
1234567
Series1 Series2