User's Manual

Table Of Contents
Page 100 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS A5E32337685
mmmmm
B: Troubleshooting
Reading erratic echo confidence weak refer to Echo Confidence (2.2.3.6.1.)
•use Auto False Echo Suppression
(AUTO ECHO SUPP) (2.2.4.3.) and
Auto Suppression Range (AUTO
SUPP RANGE) (2.2.4.4.)
use foam deflector
liquid surface vortexed decrease Fill Rate (FILL RATE/MIN)
(2.2.6.2.)
increase confidence threshold in
Echo Threshold (2.2.3.1.2.)
material filling re-locate SITRANS LR260
Reading response
slow
Fill Rate (FILL RATE/MIN)
(2.2.6.2.) setting is incorrect
increase measurement response if
possible
Reads correctly but
occasionally reads
high when vessel is
not full
detecting close range
echo
build up near top of
vessel or nozzle
nozzle problem
clean the antenna
•use Auto False Echo Suppression
(AUTO ECHO SUPP) (2.2.4.3.) and
Auto Suppression Range (AUTO
SUPP RANGE) (2.2.4.4.)
Level reading lower
than material level
material is within Near
Range zone
multiple echoes
processed
decrease Near Range (2.2.3.5.1.)
(minimum value depends on
antenna type)
raise SITRANS LR260
•ensure Algorithm (2.2.3.1.1.) is set to
tF (trueFirst echo)
vessel near empty and
low dK material
set Position (2.2.3.1.3.) to Hybrid
Symptom Cause Action (cont’d)