B: Troubleshooting
S: 34 * Factory calibration for the device
has been lost.
Repair required: contact your local
Siemens representative.
S: 35 * Factory calibration for the device
has been lost.
Repair required: contact your local
Siemens representative.
S: 36 * Unable to start microwave module. Cycle power. If fault persists, contact
your local Siemens representative.
S: 37 * Measurement hardware problem. Cycle power. If fault persists, contact
your local Siemens representative.
S: 38 * Microwave module hardware fail-
ure: unable to calculate distance
measurement.
Cycle power. If fault persists, contact
your local Siemens representative:
repair required.
S: 43 * Factory calibration for the radar
receiver has been lost.
Repair required: contact your local
Siemens representative.
S: 64
*
SPC42 Access Error Cycle power. If fault persists, contact
your local Siemens representative.