A5E32337685 SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 97
mmmmm
B: Troubleshooting
S: 15 Upper and Lower input values (Pro-
cess Value Scale) for AIFB2 are the
same.
Check configuration for AIFB2.
Ensure that Upper Value and Lower
Value (Process Value Scale) are not
the same.
S: 17 Calibration interval as defined in
Maintenance Required Limit has
expired.
Perform calibration.
S: 18 Calibration interval as defined in
Maintenance Demanded Limit has
expired.
Perform calibration.
S: 25 Internal device error. Reset power. If fault persists, con-
tact your local Siemens representa-
tive.
S: 28 * Internal device failure caused by a
RAM memory error.
Repair required: contact your local
Siemens representative.
S: 29 * EEPROM damaged. Repair required: contact your local
Siemens representative.
S: 31 * Flash error. Repair required: contact your local
Siemens representative.
S: 32
*
IDENT number conflict Ensure value of the Ident number
selector is correct for the network
configuration. If it is correct, the
device needs to be re-parameterized
by the PLC.
S: 33 * Factory calibration for the internal
temperature sensor has been lost.
Repair required: contact your local
Siemens representative.
General Fault Codes (cont’d) (cont’d)
Code /
Icon
Meaning Corrective Action