User's Manual
Table Of Contents
- Table of Contents
- SITRANS LR250 Overview
- Specifications
- Installation
- Wiring
- Operating via the handheld programmer
- Operating via SIMATIC PDM
- Functions in SIMATIC PDM
- Quick Start Wizard via SIMATIC PDM
- Changing parameter settings using SIMATIC PDM
- Parameters accessed via pull-down menus
- Operating via AMS Device Manager
- Functions in AMS Device Manager
- Features of AMS Device Manager
- Device Description (DD)
- Configuring a new device
- Startup
- Pull-down menu access
- Device configuration
- Quick Start Wizard via AMS Device Manager
- Maintenance and Diagnostics
- Remaining Device Lifetime [see Remaining Device Lifetime (4.2.) on page 99]
- Remaining Sensor Lifetime [see Remaining Sensor Lifetime (4.3.) on page 102]
- Service Schedule [see Service Schedule (4.4.) on page 105]
- Calibration Schedule [see Calibration Schedule (4.5.) on page 108]
- Electronic Temperature
- Wear (see Wear on page 56)
- Communication
- Security
- Device Diagnostics
- AMS Menu Structure
- Functions in AMS Device Manager
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Analog Output
- Maximum Process Temperature Chart
- Process Pressure/Temperature derating curves
- Loop power
- Appendix E: Application Examples
- Appendix F: HART Communications
- Appendix G: ATEX Certificates
- Appendix H: Firmware Revision History
- Glossary
- Index
- LCD menu structure
Page 82 SITRANS LR250 (HART) – INSTRUCTION MANUAL 7ML19985JE03
mmmmm
Parameters
2.4.2. Fill Rate
Defines the maximum rate at which the reported sensor value
1)
is allowed to
increase. Allows you to adjust the SITRANS LR250 response to increases in the
actual material level. Fill Rate is automatically updated whenever Response
Rate is altered.
Enter a value slightly greater than the maximum vessel-filling rate, in units per
minute.
2.4.3. Empty rate
Defines the maximum rate at which the reported sensor value
2)
is allowed to
decrease. Adjusts the SITRANS LR250 response to decreases in the actual
material level. Empty Rate is automatically updated whenever Response Rate
is altered.
Enter a value slightly greater than the vessel’s maximum emptying rate, in units
per minute.
1)
The value produced by the echo processing which represents the distance from sen-
sor reference point to the target (see Sensor Mode (2.2.2.) on page 79 for an illustra-
tion).
Options
Range: 0 to 99999 m / min.
Response Rate (2.4.1.) Fill Rate
* Slow 0.1 m/min (0.32 ft/min)
Medium 1.0 m/min (3.28 ft/min)
Fast 10.0 m/min (32.8 ft/min)
Altered by:
Response Rate (2.4.1.)
Related
parameters
Units (2.2.1.)
2)
The value produced by the echo processing which represents the distance from sen-
sor reference point to the target (see Sensor Mode (2.2.2.) on page 79 for an illustra-
tion).
Options
Range: 0 to 99999 m / min.
Response Rate (2.4.1.) Empty Rate
* Slow 0.1 m/min (0.32 ft/min)
Medium 1.0 m/min (3.28 ft/min)
Fast 10.0 m/min (32.8 ft/min)
Altered by:
Response Rate (2.4.1.)
Related
parameters
Units (2.2.1.)