Page 46 SITRANS LR250 (HART) – INSTRUCTION MANUAL 7ML19985JE03
mmmmm
SIMATIC PDM
4) In Step 4 – Ranges
a) Enter parameter values.
b) Click on Linearization.
c) In the Linearization window click on the appropriate Breakpoint tab to open the
dialog window.
d) Enter the desired level and associated volume values
1)
, and click on OK.
e) In the Step 4 window, click on NEXT.
Note: The Reset button resets values to the values in the offline table.
1)
For display purposes we recommend entering values in ascending order (that is, the low-
est values in level 1 and volume 1).
Linearization
Breakpoints 1 - 8