User's Manual

Table Of Contents
Page 144 SITRANS LR250 (HART) – INSTRUCTION MANUAL 7ML19985JE03
mmmmm
E: Application Examples
Application with Stillpipe
This application is to obtain a level measurement and corresponding 4 to 20 mA output
proportional to the oil level in a fuel storage vessel.
Notes:
Near Range (2.8.1.) (Blanking) will be set at the factory. Check the Process Device Tag for
specific values.
Suitable pipe diameters are 40 mm (1.5”) to 100 mm (4”).
The pipe diameter must be matched with the horn size. Use the largest horn size that will
fit the stillpipe/bypass pipe (see
Dimensions
on page 12).
See
Mounting on a Stillpipe or Bypass Pipe
on page 22 for installation guidelines.
Parameter
type
Parameter No./Name
Options/
Values
Function
Quick Start
Wizard
Material (1.2.) LIQUID LOW DK
Response Rate (1.3.) MED Medium =1 m/minute
Units (1.4.) M meters
Operation Mode (1.5.) LEVEL
Level is reported as Volume
when a vessel shape is
selected.
Low Calibration Point (1.6.) 5 5 m (16.5 ft)
High Calibration Point (1.7.) 0.5 0.5 m (1.64 ft)
Apply? (Apply changes) (1.8.) YES
Transfers Quick Start settings to
device.
Indepen-
dent
parameters
Propagation Factor (2.8.3.)
a)
a)
The recommended values for the propagation factor and for CLEF range are dependent
on the stillpipe diameter. See
Propagation Factor/Stillpipe Diameter
on page 145 for val-
ues.
0.990
P.F. for a 50 mm (1.96") I.D.
stillpipe
Position Detect (2.8.4.2.) HYBRID
CLEF Range (2.8.4.4.)
a
4.3 4.3 m (14.10 ft)
Low Calibration Pt. is 5 m (16.5 ft) from
the sensor reference point.
High Calibration Pt. is 0.5 m (1.65 ft)
from the sensor reference point.
The stillpipe inside diameter is
50 mm (1.96").
The maximum rate of filling or
emptying is about 0.1 m (4")/min.
5 m
0.5 m
sensor
reference
point
Low Cal.
Pt.
High
Cal.
Pt.
50 mm I.D.
vent hole