User's Manual

Table Of Contents
7ML19985JE03 SITRANS LR250 (HART) – INSTRUCTION MANUAL Page 125
mmmmm
C: Maintenance
Appendix C: Maintenance
SITRANS LR250 requires no maintenance or cleaning under normal operating conditions.
Under severe operating conditions, the horn antenna may require periodic cleaning. If
cleaning becomes necessary:
Note the antenna material and the process medium, and select a cleaning solution
that will not react adversely with either.
Remove the instrument from service and wipe the antenna clean using a cloth and
suitable cleaning solution.
Unit Repair and Excluded Liability
All changes and repairs must be done by qualified personnel, and applicable safety
regulations must be followed. Please note the following:
The user is responsible for all changes and repairs made to the device.
All new components must be provided by Siemens Milltronics Process Instruments
Inc.
Restrict repair to faulty components only.
Do not re-use faulty components.
Replacing the antenna
If the antenna requires replacement due to damage, it may be replaced without
need for re-calibration.
Changing to a different antenna type may be performed by a Siemens authorized
repair center.