7ML19985JE03 SITRANS LR250 (HART) INSTRUCTION MANUAL Page 123
mmmmm
B: Troubleshooting
Display shows
S: 0 LOE
location or aiming:
poor installation
flange not level
Auto False Echo Suppres-
sion may be incorrectly
applied
check to ensure nozzle is vertical
ensure end of antenna protrudes from
end of nozzle
•review
Auto False Echo Suppression
(2.8.7.1.)
on page 130
ensure Auto False Echo Suppression
Range is set correctly
Display shows
S: 0 LOE
antenna malfunction:
temperature too high
physical damage
excessive foam
multiple echoes
check Current Internal Temperature (3.2.1.)
use foam deflector or stillpipe
relocate
•use a defoamer
•set Algorithm (2.8.4.1.) to F (First echo)
Reading does not
change, but the level
does
SITRANS LR250 processing
wrong echo, for example, ves-
sel wall, or structural member
re-locate SITRANS LR250
check nozzle for internal burrs or welds
rotate instrument 90°
•use Auto False Echo Suppression (2.8.7.1.)
if necessary: see
Auto False Echo Sup-
pression (2.8.7.1.)
on page 130
Measurement is consis-
tently off by a constant
amount
setting for Low Calibration
Pt. (2.3.1.) not correct
setting for Sensor Offset
(2.3.3.) not correct
check distance from sensor reference
point to Low Calibration Pt. (2.3.1.)
check Sensor Offset (2.3.3.)
Screen blank
power error
check nameplate rating against voltage
supply
check power wiring or source
too much load resistance
change barrier type, or
remove something from the loop, or
increase supply voltage
Reading erratic
echo confidence weak
refer to Confidence (2.8.6.1.)
•use Auto False Echo Suppression (2.8.7.1.)
and Auto False Echo Suppression Range
(2.8.7.2.)
use foam deflector or stillpipe
liquid surface vortexed
decrease Fill Rate (2.4.2.)
relocate instrument to side pipe
increase confidence threshold in Echo
Threshold (2.8.4.3.)
material filling re-locate SITRANS LR250
Operation Troubleshooting (cont’d)
Symptom Cause Action