B: Troubleshooting
Operation Troubleshooting
Operating symptoms, probable causes, and resolutions.
S: 49
* EEPROM corrupt.
Repair required: contact your local Sie-
mens representative.
S: 50
* EEPROM corrupt.
Repair required: contact your local Sie-
mens representative.
S: 51
* EEPROM corrupt.
Repair required: contact your local Sie-
mens representative.
S: 52
Fail-safe is activated. Possible causes:
1) hardware failure
2) memory failure
3) Fail-safe LOE timer expired– possible
causes: faulty installation, antenna mate-
rial buildup, foaming/other adverse process
conditions, invalid calibration range.
For 3):
Correct configuration; ensure installation
is correct;
no antenna buildup;
adjust process conditions to minimize
foaming/other adverse conditions;
correct calibration range.
If fault persists, or for 1) and 2), contact
your local Siemens representative.
S: 53
*
Configuration lost: one or more parameter
settings have been lost. This may occur
after a firmware upgrade causes user
parameters to be reset.
Restore user parameters using SIMATIC
PDM.
Operation Troubleshooting
Symptom Cause Action
Display shows
S: 0 LOE
level or target is out of range
check specifications
check Low Calibration Pt. (2.3.1.)
increase Confidence (2.8.6.1.)
Display shows
S: 0 LOE
material build-up on antenna
clean the antenna
re-locate SITRANS LR250