User's Manual
Table Of Contents
- Table of Contents
- SITRANS LR250 Overview
- Specifications
- Installation
- Wiring
- Operating via the handheld programmer
- Operating via SIMATIC PDM
- Functions in SIMATIC PDM
- Quick Start Wizard via SIMATIC PDM
- Changing parameter settings using SIMATIC PDM
- Parameters accessed via pull-down menus
- Operating via AMS Device Manager
- Functions in AMS Device Manager
- Features of AMS Device Manager
- Device Description (DD)
- Configuring a new device
- Startup
- Pull-down menu access
- Device configuration
- Quick Start Wizard via AMS Device Manager
- Maintenance and Diagnostics
- Remaining Device Lifetime [see Remaining Device Lifetime (4.2.) on page 99]
- Remaining Sensor Lifetime [see Remaining Sensor Lifetime (4.3.) on page 102]
- Service Schedule [see Service Schedule (4.4.) on page 105]
- Calibration Schedule [see Calibration Schedule (4.5.) on page 108]
- Electronic Temperature
- Wear (see Wear on page 56)
- Communication
- Security
- Device Diagnostics
- AMS Menu Structure
- Functions in AMS Device Manager
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Analog Output
- Maximum Process Temperature Chart
- Process Pressure/Temperature derating curves
- Loop power
- Appendix E: Application Examples
- Appendix F: HART Communications
- Appendix G: ATEX Certificates
- Appendix H: Firmware Revision History
- Glossary
- Index
- LCD menu structure
7ML19985JE03 SITRANS LR250 (HART) – INSTRUCTION MANUAL Page 121
mmmmm
B: Troubleshooting
S: 29
* EEPROM damaged.
Repair required: contact your local Sie-
mens representative.
S: 31
*Flash error.
Repair required: contact your local Sie-
mens representative.
S: 33
*
Factory calibration for the internal temper-
ature sensor has been lost.
Repair required: contact your local Sie-
mens representative.
S: 34
*
Factory calibration for the device has been
lost.
Repair required: contact your local Sie-
mens representative.
S: 35
*
Factory calibration for the device has been
lost.
Repair required: contact your local Sie-
mens representative.
S: 36
* Unable to start microwave module.
Cycle power. If fault persists, contact your
local Siemens representative.
S: 37
* Measurement hardware problem.
Cycle power. If fault persists, contact your
local Siemens representative.
S: 38
* Failure in the device electronics.
Cycle power. If fault persists, contact your
local Siemens representative: repair
required.
S: 43
*
Factory calibration for the radar receiver
has been lost.
Repair required: contact your local Sie-
mens representative.
S: 45
*
No valid boot program detected: firmware
corrupt.
Repair required: contact your local Sie-
mens representative.
S: 48
*
User configuration is invalid. One or more
of parameters: Low Calibration Point, High
Calibration Point, Volume breakpoints, and/
or Auto False-Echo Suppression, are set to
invalid values.
• Reconfigure the unit.
• Ensure the difference between High Cali-
bration Point and Low Calibration Point is
not less than or equal to zero; do a Mas-
ter Reset.
General Fault Codes (cont’d)
Code /
Icon
Meaning Corrective Action