User's Manual

Table Of Contents
Parameter reference
SITRANS LR250 (HART)
160 Operating Instructions, 01/2014, A5E32220602-AB
Reminder 2 before Calibration (Demanded) (4.5.6.)
If
Time Until Next Calibration (4.5.3.)
is equal to or less than this value, the device generates a
Maintenance Demanded
reminder.
Values
Range: 0 to Calibration Interval (4.5.1.)
Default: 0.164 years
1. Modify values as required.
2. Set Activation of Reminders (4.5.4.) to the desired option.
Maintenance Status (4.5.7.)
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu
View Device Status
, click on the
Maintenance
tab and check
the
Calibration Schedule Status
window.
Acknowledge Status (4.5.8.)
Indicates which level of maintenance reminder has been acknowledged.
In SIMATIC PDM, open the menu
View Device Status
, click on the
Maintenance
tab and check
the
Calibration Schedule Status
window.
Acknowledge (4.5.9.)
Acknowledges the current maintenance reminder.
To acknowledge a reminder via SIMATIC PDM:
1. Open the menu
View Device Status
and click on the
Maintenance
tab.
2. In the
Service Schedule Status
section, click on
Acknowledge Warnings
.
To acknowledge a reminder via the handheld programmer:
1. Press
RIGHT arrow
twice to open parameter view and activate
Edit
Mode.
2. Press
RIGHT arrow
to acknowledge the reminder.
Manufacture Date (4.6.)
Read only. The date of manufacture of the SITRANS LR250 (yy mm dd).