Parameter reference
SITRANS LR250 (HART)
Operating Instructions, 01/2014, A5E32220602-AB
159
Time Until Next Calibration (4.5.3.)
Read only.
Calibration Interval (4.5.1.)
less
Time Since Last Calibration (4.5.2.)
.
Activation of Reminders (4.5.4.)
Note
To modify this parameter v
ia SIMATIC PDM it must be accessed via the pull-down menu
Device
Maintenance.
Allows you to enable a maintenance reminder.
Options
Timer OFF
ON - no reminders checked
ON - Reminder 1 (Maintenance Required) checked
*
ON - Reminders 1 and 2 checked
ONReminder 2 (Maintenance Demanded) checked
1. First set the values in
Reminder 1 before Calibration (Required) (4.5.5.)
/
Reminder 2 before
Calibration (Demanded) (4.5.6.)
.
2. Select the desired
Activation of Reminders
option.
Reminder 1 before Calibration (Required) (4.5.5.)
If
Time Until Next Calibration (4.5.3.)
is equal to or less than this value, the device generates a
Maintenance Required
reminder.
Values
Range: 0 to Calibration Interval (4.5.1.)
Default: 0.164 years
1. Modify values as required.
2. Set
Activation of Reminders (4.5.4.)
to the desired option.