Parameter reference
SITRANS LR250 (HART)
Operating Instructions, 01/2014, A5E32220602-AB
151
Time in Operation (4.3.2.)
The amount of time the sensor has been operating. Can be reset to zero after performing a
service or replacing the sensor.
To reset to zero:
In SIMATIC PDM, open the menu
Device Maintenance
, click on the
Remaining Sensor
Lifetime
tab, and click on
Sensor Replaced
to restart the timer and clear any fault messages.
Via the handheld programmer, manually reset
Time in Operation (4.3.2.)
to zero.
Remaining Lifetime (4.3.3.)
Read only.
Lifetime (Expected) (4.3.1.)
less
Time in Operation (4.3.2.)
.
Activation of Reminders (4.3.4.)
Note
To modify this parameter via SIMATIC PDM it must be accessed via the pull
-down menu
Device Maintenance
.
Allows you to enable a maintenance reminder.
Options
Reminder 1 (Maintenance Required)
Reminder 2 (Maintenance Demanded)
Reminders 1 and 2
*
OFF
1. First set the values in
Reminder 1 before Lifetime (Required) (4.3.5.)
/
Reminder 2 before
Lifetime (Demanded) (4.3.6.)
.
2. Select the desired
Activation of Reminders
option.
Reminder 1 before Lifetime (Required) (4.3.5.)
If
Remaining Lifetime (4.3.3.)
is equal to or less than this value, the device generates a
Maintenance Required
reminder.
Values
Range: 0 to Lifetime (Expected) (4.3.1.)
Default: 0.164 years
1. Modify values as required.
2. Set
Activation of Reminders (4.3.4.)
to the desired option.