User's Manual

Table Of Contents
Parameter reference
SITRANS LR250 (HART)
140 Operating Instructions, 01/2014, A5E32220602-AB
Hover Level (2.8.7.3.)
Note
Changes take effect only at the next measurement.
Defines how high the TVT (Time Varying Threshold) is placed above the noise floor of the echo
profile, as a percentage of the difference between the peak of the largest echo in the profile and
the noise floor. See
Auto False Echo Suppression (2.8.7.1.)
for an illustration.
Values
Range: 0 to 100%
Default: 40%
When the device is located in the center of the vessel, the TVT hover level may be lowered to
increase the confidence level of the largest echo.
Shaper Mode (2.8.7.4.)
Enables/disables the TVT shaper.
Options
ON
*
OFF