Parameter reference
SITRANS LR250 (HART)
Operating Instructions, 01/2014, A5E32220602-AB
137
Auto False Echo Suppression (2.8.7.1.)
Used together with
Auto False Echo Suppression Range (2.8.7.2.)
to screen out false echoes in
a vessel with known obstructions. A ’learned TVT’ (time varying threshold) replaces the default
TVT over a specified range. See
Auto False Echo Suppression (Page 221)
for a more detailed
explanation.
Note
Make sure material level is below all known obstructions when Auto False Echo
Suppression is used to learn the echo profile. (An empty or almost empty vessel is
recommended.)
Note the distance to material level when Auto False Echo learns the environment. Set
Auto False Echo Suppression Range to a shorter distance to avoid the material echo
being screened out.
Set Auto False Echo Suppression and Auto False Echo Suppression Range during
startup, if possible.
If the vessel contains an agitator it should be running.
Before adjusting these parameters, rotate the instrument for best signal (lower false-
echo
amplitude).
Before Auto False Echo Suppression
TVT Hover Level
echo marker
default TVT
material level
false echo